O-Ring Care; Metal Seal Care - Agilent Technologies Helium G8610 Series Field Installation Instructions

Ion source replacement for leak detection
Hide thumbs Also See for Helium G8610 Series:
Table of Contents

Advertisement

Vacuum Products Division

1.2.3 O-ring Care

When removing, checking, or replacing O-rings, keep in mind the following:
NOTE
CAUTION

1.2.4 Metal Seal Care

CAUTION
Manual 699910005, Hot Ion Source Replacement
Agilent recommends replacing all O-rings during routine maintenance or
during any maintenance procedure requiring that O-rings be removed.
Remove O-rings carefully with your fingers. Do not use metal tools for
this task; this prevents scratching of any sealing surfaces.
Wipe all O-rings clean with a lint-free cloth before installation to
ensure that no foreign matter is present to impair the seal.
Do not use grease or any other substance on O-rings that will
come in contact with the vacuum surfaces.
Do not use alcohol, methanol or other solvents on O-rings. Doing
so causes deterioration and reduces their ability to hold a
vacuum.
Agilent does not recommend the use of vacuum grease. If
applicable, apply a small amount of Apiezon® L grease and wipe
the O-rings shiny dry.
Metal Seals must be replaced any time a spectrometer is opened. All
fasteners must be installed and torqued per assembly procedure
specifications. Remove Metal Seals carefully with your fingers or a soft
tool. Metal tools scratch sealing surfaces.
Metal Seals are supplied in pre-cleaned condition. No cleaning is
required. If necessary, Metal Seals can be cleaned using the
recommended solvents. Wipe Metal Seals clean with a lint-free
cloth before installation to ensure that no foreign matter impairs
the seal.
Do not use grease or any other substance on Metal Seals that
will come in contact with the spectrometer.