Spanner/wrench
An equipment error has
shows on the
occurred.
display.
Thermometer
Ventilation slits blocked.
shows in the display
(ATMOS S 351 OT
has overheated).
Ambient temperature
too high.
Fan is defective.
Only a low vacuum
A low target vacuum level
level can be set with
has been set with the
the foot controller.
buttons.
If the symbol disappears, the device
was able to correct the error. Nev-
ertheless, have your suction device
checked by ATMOS Service.
Only emergency operation is possi-
ble as long as the symbol shows. Call
ATMOS Service.
Check the ventilation slits (bottom
of the device); they must not be
blocked.
Use the ATMOS S 351 OT only in the
specified temperature range. Try
using auto standby (less heat build-
up).
Call ATMOS Service.
Set the target vacuum to a higher
value (or MAX) using the + button to
obtain a larger control range for the
foot controller.
Troubleshooting
39