4. Firmware Update

Requirements

Computer with USB2.0 or higher
USB drivers installed
HyperTerminal (or other terminal emulation program)

Firmware Update Procedure

1. Connect the USB cable from computer to the powered off AEMC 6292.
2. Launch HyperTerminal.
3. Configure HyperTerminal Com Port as shown below (same as Control Panel > Device Manager).
57600 bps, 8 bits, No parity, No Flow Control
4. Press and hold the ZERO and TWO keys.
5. Power on the instrument, while pressing the ZERO and TWO keys.
6. The following screen will be displayed.
v1 - 03/15
SERVICE/CALIBRATION MANUAL
FIRMWARE UPDATE!
- 8 -