13.Technical Specifications

Type:
Non Contact Thermometer NC 100
Measurement
Body mode: 34.0-42.2 °C / 93.2-108.0 °F
range:
Object mode: 0-100.0 °C / 32-212.0 °F
Resolution: 0.1 °C / °F
Measurement
Laboratory:
0.2 °C, 36.0  39.0 °C
accuracy:
0.4 °F, 96.8  102.2 °F
Display:
Liquid Crystal Display, 4 digits plus special
icons
Acoustic:
The unit is turned ON and ready for the
measurement: 1 short beep.
Complete the measurement: 1 long beep
(1 sec.) if the reading is less than 37.5 °C /
99.5 °F, 10 short «beep» sounds, if the
reading is equal to or greater than 37.5 °C
/ 99.5 °F.
System error or malfunction: 3 short «bi»
sounds.
NC 100
Memory:
Auto-Display the last measured temperature
30 readings recall in the memory mode
Backlight:
The display light will be GREEN for 3
seconds, when the unit is turned ON.
The display light will be GREEN for 5
seconds, when a measurement is completed
with a reading less than 37.5 °C / 99.5 °F.
The display light will be RED for 5 seconds,
when a measurement is completed with a
reading equal to or higher than 37.5 °C /
99.5 °F.
Operating
Body mode: 15-40.0 °C / 59-104.0 °F
conditions:
Object mode: 5-40.0 °C / 41-104.0 °F
Storage
-25 - +55 °C / -13 - +131 °F
conditions:
15 - 95 % relative maximum humidity
Automatic
Approx. 1 minute after last measurement
Switch-off:
has been taken.
Battery:
2 x 1.5 V alkaline batteries; size AAA
Battery
approx. 2000 measurements
lifetime:
(using new batteries)
Dimensions: 150 x 40 x 39 mm
Weight:
81 g (with batteries), 56 g (w/o batteries)
IP Class:
IP21
Reference to
ASTM E1965;
standards:
IEC 60601-1; IEC 60601-1-2 (EMC);
IEC 60601-1-11
Expected
service life: 5 years or 12000 measurements
This device complies with the requirements of the Medical
Device Directive 93/42/EEC.
Technical alterations reserved.
According to the Medical Product User Act a biennial tech-
nical inspection is recommended for professional users.
Please observe the applicable disposal regulations.

14.www.microlife.com

Detailed user information about our thermometers and
blood pressure monitors as well as services can be found
at www.microlife.com.
7
EN

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents