Advertisement
To turn on the thermometer, push the button on the top; a short signalling tone [beep] signals «thermometer ON.» A display test is performed. Then at an ambient temperature of less than 32°C, an «L» and a flashing «°C» appear at the upper right of the display. The thermometer is now ready for use.
During a measurement, the current temperature is displayed continuously. The «°C» symbol flashes during this period of time. If the signalling tone sounds (beep-beep-beep-beep 10 times) and the «°C» is no longer flashing, this means that the measured increase in temperature is now less than 0.1°C in 16 seconds and that the thermometer is ready for reading.
To prolong the life of the battery, turn it off after use by pressing the operating button briefly. Otherwise the thermometer will automatically turn off after about 10 minutes.
If you hold the button down for more than 3 seconds when turning the thermometer on, the maximum temperature stored automatically at the last measurement will be displayed. At the same time, a small «M» for memory will appear at the right of the display. About 2 seconds after release the button, this value will disappear and the thermometer will be ready for measurement.
Position the thermometer in one of the two pockets under the tongue, to the left or right of the root of the tongue. The measuring sensor must be in good contact with the tissue. Close your mouth and breath evenly through the nose to prevent the measurement from being influenced by inhaled/ exhaled air. Approx. measuring time: 1 minute!
This is the most reliable measuring method, and is especially suitable for infants and small children. Carefully insert the measuring sensor of the thermometer 2 to 3 cm into the anal aperture. Approx. measuring time: 1 minute!
To receive more reliable results we recommend to measure temperature orally or rectally. Approx. measuring time: 5 to 10 minutes!
Name: Isopropyl alcohol 70%; immerse: max. 24 hours.
When the «» symbol (upside-down triangle) appears in the lower right of the display, the battery is exhausted and thermometer needs to be replaced. For battery disposal please refer to «safety instructions».
Type | Maximum thermometer |
Measurement range | 32.0°C to 42.9°C Temp. < 32.0 °C:display «L» for low (too low) Temp. > 42.9°C:display «H» for high (too high) |
Measurement accuracy | ± 0.1°C between 34°C and 42°C at an ambient temperature of 18°C to 28°C |
Self-test: | Automatic internal check at a test value of 37°C; if there is a deviation of > 0.1°C, «ERR» (error) is displayed |
Display | Liquid crystal display (LCD) with 3 digits; Smallest unit of display: 0.1°C |
Signalling tone | For signalling that the thermometer is ready to use and that the temperature increase is less than 0.1°C / 16 seconds |
Memory | For storing the last measured value |
Storage temperature | –10°C to +60°C |
Battery | IEC type SR41 Silver oxide high drain |
Reference to standards | EN12470-3: 2000, Clinical Thermometers, ASTM standard |