Condumax II Transportable
Sampling System User's Manual
Appendix B
Hazardous Area Certifi cation
The Condumax is certifi ed compliant to the ATEX Directive (94/9/EC) and IECEx for use
within Zone 1 and Zone 2 Hazardous Areas and has been assessed so by TRaC (Notifi ed
Body 0891).
The Condumax is certifi ed compliant to North American Standards (USA and Canada)
for use within Class I, Division 1, Groups B, C and D Hazardous Locations and has been
assessed so by CSA.
B.1
Product Standards
This product conforms to the Standards:
EN60079-0:2012
EN60079-1:2007
CSA C22.2 No. 142-M1987
C22.2 No. 30-M1986
B.2
Product Certifi cation
This product is attributed with the product certifi cation codes:
ATEX & IECEx
II 2G Ex d IIB + H2 T4 Gb (-40°C to +45°C)
North American
Class I, Division 1, Groups B, C & D, T4, Tamb -40°C to +60°C
B.3
Global Certifi cates/Approvals
ATEX
IECEx
cCSAus
TC TR Ex
These certifi cates can be viewed or downloaded from our website at:
http://www.michell.com
Michell Instruments
TRAC11ATEX21319X
TRC 11.0008X
1701657
RU C-GB. ГБ05.B.00152
Special attention should be paid to the Special Conditions
for Safe Use and the Conditions of Certifi cation listed in the
certifi cates shown on the website.
IEC60079-0:2011
IEC60079-1:2007
UL508
UL1203
T3 Gb (-40°C to +60°C)
APPENDIX B