Troubleshooting Guide - Acon On Call ExtraMobile User Manual

Blood glucose monitoring system
Table of Contents

Advertisement

The meter has built-in messages to alert you of problems. When error messages
appear, note the error number. Turn off the meter and then follow these instructions.
Display
Battery may be damaged or
empty
Meter fails
to turn on
Meter is too cold
Power On self check error
Internal calibration check error
Test strip was removed during
the test
Sample was applied to the
test strip too soon
Test strip is contaminated or
used
Sample was applied to the
test strip too soon
Insufficient sample
Sample application error due
to late sample re-dosing

Troubleshooting Guide

Causes
Replace the battery.
If the meter has been exposed to or stored
in cold conditions, wait 30 minutes. Allow
the meter to reach room temperature then
repeat the test.
Remove the battery for 30 seconds and
then put the battery back and turn meter
on again.
If the problem persists, please contact
your local distributor.
If a cell phone, radio frequency source or
a high power electrical source is nearby,
place more distance between the meter
and any of these sources then retest.
If the problem persists, please contact
your local distributor.
Repeat the test and ensure that the test
strip remains in place.
Repeat the test and apply the sample after
the blood drop/test strip symbol appears.
Repeat the test with a new test strip.
Repeat the test and apply the sample the
after blood drop/test strip symbol appears.
Repeat the test and apply enough sample
to fill the test strip check window.
Repeat the test and apply enough sample
to fill the test strip check window within 3
seconds.
35
Solution