front panel will illuminate momentarily and an audible signal will briefly
sound confirming that the LEDs and audible signal are functioning
properly. The unit will operate in "start up" mode with the Low Oxygen light
lit until oxygen stabilization is achieved, at which time the Normal Oxygen
light will come on and remain lit. The "start up" mode may take up to 15
minutes.
NOTE– DeVilbiss recommends for optimal service life that the Drive
DeVilbiss Oxygen Concentrator to be operated for at least 30 minutes
after it is powered on. Shorter periods of operation, operating in extreme
temperature/humidity conditions or in the presence of contaminates, and/
or handling and storage conditions outside those specified, may affect the
long term reliable operation of the product.
4.
Slowly turn the flow meter knob until the flow meter ball is centered on the
line next to the appropriate flow rate.
NOTE– When the flow meter knob is turned clockwise, the flow
decreases (and eventually will shut off the oxygen flow). When the knob is
turned counter-clockwise, the flow increases.
NOTE– For prescriptions of 10 LPM, be sure the ball is centered on
the 10 liter line; the ball should not touch the red line. Setting the flow
higher than 10 may cause the oxygen purity level to drop.
NOTE– The low-flow alarm may activate if the flow meter ball is set
below 2 lpm. The unit will continue to run; however, the Service Required
light will come on accompanied by an audible alarm. Adjust the flow meter
to your prescribed flow.
NOTE– Do not use a low-output flow meter with this concentrator.
NOTE– The unit may require up to 15 minutes for the oxygen
concentration and flow rate to stabilize. The flow rate should be monitored
and readjusted if necessary.
5.
The flow meter has a locking device. If it is necessary to preset and lock
in the prescribed flow rate, tighten the set screw located on the hex nut
just below the control knob using a 1/16" Allen bit. No adjustment can be
made without loosening the set screw.
6.
The Drive DeVilbiss oxygen concentrator is now ready for use.