SYSTEM OPERATION

The Drive DeVilbiss Oxygen Concentrator uses a pressure swing adsorption
system. The air is drawn into the unit through air filters and into a double-head
compressor.
A pneumatic diagram of the system is shown on page 29.
The compressed air passes through a rotary valve, which is cycled at a pre-
determined rate, and is directed into one of two sieve beds. The sieve beds
contain molecular sieve material which is a synthetically-produced inorganic
silicate. It is very porous and has the unique ability to selectively adsorb nitrogen
from the air as it passes through the sieve bed.
As one bed is being pressurized, the other bed is quickly depressurized. This
allows the nitrogen that was adsorbed during its pressurization cycle to be
exhausted from the sieve material.
The nitrogen is released through an exhaust port located on the rotary valve
assembly. The port is connected to the exhaust muffler.
Also during each bed pressurization, a small amount of oxygen flows through an
orifice from the pressurized bed into the depressurizing bed. This helps purge the
nitrogen from the depressurizing bed.
The beds will continue to be alternately pressurized and depressurized as the
unit operates.
Oxygen leaving the sieve beds is directed through a check valve to the
accumulator tank. A pressure regulator on the tank controls the oxygen pressure
as it leaves the accumulator and enters the flow meter. The flow meter allows the
oxygen flow to be controlled and adjusted to the level prescribed by the patient's
physician. From the flow meter the oxygen passes through the final bacteria filter
and finally the oxygen outlet port to the patient.
The Drive DeVilbiss Oxygen Concentrator operates on a timed cycle (2.5 sec. @
flows > 4.5 LPM) that is controlled by the PC board. The PC board will send
voltage to the valve causing it to shift and alternately pressurize the sieve beds.
The PC board also activates the electronic alert system. Low flow, system
abnormality, and power failure are indicated by audible and visible alerts. A high
pressure condition will be indicated with a "popping" type sound produced by
release of pressure from a pressure relief valve on the compressor head.
The 1025 operating system incorporates "turn-down" technology. The PC board
constantly monitors the flow rate and will decrease the cycle time whenever the
flow rate is equal to or less than 4.5 LPM. Therefore it "turns-down" the cycle
based on lower oxygen demand. As a result, the unit runs cooler and less power
is consumed.
LT-2329
T R O U B L E S H O O T I N G

NORMAL OPERATING SEQUENCE

When the concentrator is turned "On," the following cycling sequence can be
observed by attaching a pressure gauge to the accumulator tank test point.
1.
2.
3.
4.
5.
6.
than 2 seconds.
12
The rotary valve is quickly cycled several times to relieve residual bed
pressure preventing a static condition in the compressor. This rapid cycling
only happens on start-up and is clearly heard as pressure is being quickly
exhausted several times. The pressure exhausts through an exhaust
muffler that is connected to the valve.
The PC board applies a short DC voltage signal to the valve. The valve will
stop for several seconds causing the right bed to pressurize first while the
left bed depressurizes.
Voltage is again applied to the valve for a short time. The valve will stop for
approximately a second. During this time the sieve bed pressures are
equalized.
A short DC voltage signal is again applied to the valve. The valve will stop
for several seconds causing the left bed to pressurize while the right bed
depressurizes.
A short DC voltage signal is again applied to the valve. The valve will stop
for approximately a second. During this time, the sieve bed pressures are
equalized.
The cycle then repeats with step 2.