DeVilbiss Drive 1025DS Service Manual page 5

10-liter oxygen concentrator
Table of Contents

Advertisement

WARNING
MR Unsafe
Do not bring the device or accessories into a Magnetic Resonance (MR) environment as it may cause unacceptable risk to the patient or damage to
the oxygen concentrator or MR medical devices. The device and accessories have not been evaluated for safety in an MR environment.
Do not use the device or accessories in an environment with electromagnetic equipment such as CT scanners, Diathermy, RFID and electromagnetic
security systems (metal detectors) as it may cause unacceptable risk to the patient or damage to the oxygen concentrator. Some electromagnetic
sources may not be apparent, if you notice any unexplained changes in the performance of this device, if it is making unusual or harsh sounds,
disconnect the power cord and discontinue use. Contact your home care provider.
This device is suitable for use in home and healthcare environments except for near active HF SURGICAL EQUIPMENT and the RF shielded room of
an ME SYSTEM for magnetic resonance imaging, where the intensity of Electromagnetic DISTURBANCES is high.
Use of this equipment adjacent to or stacked with other equipment should be avoided because it could result in improper operation. If such use is
necessary, this equipment and the other equipment should be observed to verify that they are operating normally.
Portable RF communications equipment (including peripherals such as antenna cables and external antennas) should be used no closer than 30 cm
(12 inches) to any part of the oxygen concentrator, including cables specified by the manufacturer. Otherwise, degradation of the performance of this
equipment could result.
CAUTION
It is very important to follow your oxygen prescription. Do not increase or decrease the flow of oxygen – consult your physician.
To prevent product damage, do not attempt to operate the unit without the air filter, or while the filter is still damp.
The surface temperature of the exhaust vents on the bottom of the unit may exceed 105.8˚F (41˚C) under certain conditions.
When device is used under extreme operating conditions, the temperature near the exhaust vents on the bottom of the unit may reach 138.2°F (59°C).
Keep body parts a minimum of 30" (76.2 cm) away from this area.
Use of harsh chemicals (including alcohol) is not recommended. If bactericidal cleaning is required, a non-alcohol based product should be used to
avoid inadvertent damage.
IMPORTANT
It is recommended that the homecare provider lock the flow control knob to prevent inadvertent adjustment. A flow setting other than prescribed may
affect the patient therapy.
Do not service or clean this device while in use with a Patient.
Do not use a low-output flow meter with this concentrator.
The Device is classified as IP21 which means it is protected against finger access to hazardous parts and protected against vertically falling water
drops.
Equipment not suitable for use in the presence of a flammable anesthetic mixture with air or with oxygen or nitrous oxide.
This device contains electrical and/or electronic equipment. Follow local governing ordinances and recycling plans regarding disposal of device
components.
SAVE THESE INSTRUCTIONS.