Menu Settings - TA Instruments DHR Series Getting Started Manual

Asphalt submersion cell
Hide thumbs Also See for DHR Series:
Table of Contents

Advertisement

NOTE: Please note that if you enter the values as prescribed below and then for any reason go back and
change any one parameter, all the previous PID settings will be altered. This applies to parameters in the
Control Section only.
Safety temperature

Menu Settings

Select the following menus settings using the keys as instructed above:
Pump: Adjustable Pump Capacity
Pump Speed: 2
Control: Intelligent Cascade Control Parameters
Control: External (Note: must have pt100 cable connected between bath and circulator. If not display
will default to Internal Control.)
Self-Tuning
Off
Cospeed
0
Control
PID values
XP
0.6
TN
70
TV
7
XPU
4.5 Default
Config: Configuration of the Circulator
Set: Serial
A-start: off
Off Mode: Pump off
Reset: no (to restore defaults select yes and then enter)
Actvar: control
Time/Date: Setting time and date
DHR/AR Series Asphalt Submersion Cell Getting Started Guide
110v/120v
Off
0
PID values
0.5
90
24
4.5 Default
Pump Start/Stop
Enter
220/240v
Page 20

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Ar series

Table of Contents