Michell Instruments Optidew User Manual page 49

High performance optical dew-point transmitter
Hide thumbs Also See for Optidew:
Table of Contents

Advertisement

Optidew Series User's Manual
Appendix C
Troubleshooting - Common Faults
A list of common faults can be found in this section, along with associated diagnostic
and troubleshooting actions.
Cannot connect to HyperTerminal / OptiSoft
Dew-point reading -100 or +473
Dew-point reading shows fi xed or ambient temperature
Fixed 100% Mirror Condition
Fixed fl ashing 0% (low) Mirror Condition
Peltier power cooling 100%
Temperature reading -100 or +473 or does not change
Dew-point reading in error
Dew-point reading unstable
The troubleshooting steps for each fault should be followed in the order that they
appear.
!
Symptom: Cannot connect to HyperTerminal / Optisoft
Cause
Resolution
Cause
Diagnosis
Resolution
Cause
Resolution Restart Optidew and restart host PC / equipment
Michell Instruments
Do not attempt to disassemble the instrument or sensor.
This will invalidate the calibration and may cause permanent
damage to the instrument, voiding the warranty.
It is permitted to remove the lid of the Optidew for the purpose
of adjusting the mirror condition potentiometer only.
The Optidew Vision must NOT be disassembled.
If unit has display it is not in REMOTE MODE
Enable REMOTE MODE by holding display button for approximately 7
seconds
Unit is RS485 / RS422 communications version
Confi rm unit is RS232 version
To connect to a standard RS232 port on a PC an appropriate RS232 to
RS422 or RS485 adaptor is required
Serial ports have stopped responding on host PC
WARNING:
APPENDIX C
41

Hide quick links:

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Optidew and is the answer not in the manual?

This manual is also suitable for:

Optidew vision

Table of Contents