Safety; Draining The Water Reservoir - Interacoustics Air Fx Instructions For Use Manual

Caloric irrigator
Hide thumbs Also See for Air Fx:
Table of Contents

Advertisement

Air Fx Irrigator Instructions for Use - US
Once the final irrigations have been completed, remove the speculum from the ear tip of the handle
and clean the handle tip with Sani-Cloth wipes.

3.6 Safety

The Air Fx air irrigator uses
heater element. While the temperature of the coolant is not displayed, it is monitored. If the coolant
temperature goes over a certain limit, the unit will stop operation and an "Over Temp Error" message is
displayed. This error could be a result of insufficient water in the water reservoir or bubbles in the water
line. Operating the irrigator in very cold (<10°C conditions can also cause this error). Make sure the
irrigator has warmed up slowly to room temperature before operating.
Over Temp Error
Call TechSupport
The air delivery temperature is also monitored for safety. If this temperature exceeds 50°C, the irrigator
will stop operation and an "Air Temp Error" message will be displayed.
Air Temp Error
Call TechSupport
If either of these conditions occurs, then shut down the irrigator and isolate the problem (ambient
temperature too low, bubbles in the water line,
Contact Interacoustics' Service Department or your local service representative if the problem cannot be
resolved.
Set the Air Fx irrigator near the edge of the table or cart in order for the hose to hang down, allowing the
handle to rest properly in its cradle. When the irrigator is not in use, coil the Air Fx handle tubing using
the hose management tape while still allowing the handle to be secured in the cradle.

3.7 Draining the Water Reservoir

The Air Fx should have the water reservoir on the top of the irrigator filled with
water
. If the
distilled or demineralized water
adding new
distilled or demineralized
following instructions. Do not store irrigator in a location below 0°C (32°F) .
1. Unscrew and remove the water reservoir cap on the top of the irrigator.
2. Connect the supplied plastic syringe and tubing to the Fill / Drain Water Outlet on the back of the
irrigator. Push the syringe plastic tubing fully into the Fill / Drain Water Outlet connector, and then
gently pull on the tubing to verify the connection is secure and will not leak.
3. Switch the Fill / Drain Valve to the Fill / Drain position.
4. Draw back the syringe to draw the
water reservoir.
5. Set the Fill / Drain Valve to the Run position.
distilled or demineralized water
distilled or demineralized water
needs to be drained prior to shipment or flushed before
water, then the Air Fx should be drained according to the
distilled or demineralized water
to maintain the temperature in the cooler /
level too low etc.).
distilled or demineralized
out of the internal tubing and
Page
16

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents