3.3 Select Automatic Control of the Irrigator

If you have an Interacoustics VN415/VO425 video recording system, then you can set up the
Interacoustics software to automatically control the IA Air Fx irrigator. In the configure test menu for
caloric testing, select the stimulus tab and then select Air Fx control and choose your temperatures. In
this menu you can also choose a test time in seconds.
The air temperature default is set to 30°C for cold irrigation and 44°C for warm irrigation. The irrigation time
is set for 60 seconds.