3.4 Select the Irrigation Temperature

Select the irrigation you wish to perform from the Interacoustics software by first clicking on the
caloric test icon
and then selecting the green arrow
to start/stop a test a caloric test.
From the pop-up menu click the button for the irrigation you would like to start and then press ok
to begin the test process.
The Irrigator will be automatically initiated to the correct temperature setting (cool or warm default
temperature). A pop-up menu will show you the status of the irrigator preparing the desired
temperature.
When the irrigator reaches the desired temperature, an indicator tone will sound and a dialog box
will appear letting you know you can now begin testing.
You may now lift the irrigator handle and position the irrigator tip in the ear canal. Once the tip is
properly positioned and the patient has been instructed, press the start button on the irrigator
handle and the caloric test recording time will start. After the designated time is reached another
indicator tone will sound and the air flow will stop. The recording will continue until the selected
caloric test completion time is reached. The elapsed test time will appear in the software.