Aqua Stim Water Irrigator Instructions for Use - US

3.4 Select the Irrigation Temperature

Select the irrigation you wish to perform from the Interacoustics software by first clicking on the caloric test
icon
and then selecting the green arrow
click the button for the test you would like to start and then press OK to begin the test process.
The Irrigator will be automatically initiated to the correct temperature setting (cool or warm default
temperature). A popup menu will show you the status of the irrigator preparation of the temperature.
When the irrigator reaches the correct temperature, you will hear a beep and a popup menu will appear
letting you know you can now begin testing.
You may now lift the irrigator handle and position the irrigator tip in the ear canal. Once the tip is properly
positioned and the patient has been instructed, press the start button on the irrigator handle and water will
begin to flow, and the test recording time will start. After the designated time is reached, the water flow will
stop and the recording will continue until the test completion time is reached.
Note: if you manually change the temperature on the Aqua Stim itself by depressing one of the
temperature buttons, it will not change the temperature in the Interacoustics software. If you cancel
the test at anytime the irrigator will go into rest mode until you click on the green arrow button to
start another test.
If you attempt to manually change the temperature once you have collected data you will receive an error
message warning you that the temperature change is not allowed.
to start/stop a caloric test. From the pop-up menu
Page
13