11 Troubleshooting

Press and hold (min. 3 seconds) the "On/off" button to reset the device
(restart/reset function).

11.1 Preparations

Disconnect all leads and pressure accessories/connections from the pressure test
kit.
WARNING
Risk of serious injury from pressurized parts!
Discharge all pressure accessories and connections before starting.
Risk of injury from live parts!
Disconnect the power supply and charging cable.
Short-circuits in the battery and its connectors can result in damage and
injury.

11.2 Notifications/malfunctions on the display

11.2.1 Display of notification/malfunctions

Notifications/faults are displayed in a separate dialog as a four-digit code; these
must be confirmed with the Ok button.
In addition to the fault code, an info text is displayed for interpreting the fault.
Furthermore, all fault messages are also recorded in the log folder in the
corresponding log files.

11.3 Display without function

If the battery of the PMS3000 was deep discharged, the battery must be charged
for at least 10 minutes before the PMS3000 can be switched back on again.
Jul-2020