BE-VIF I/F Check
Checks communications (I2C bus and SPI bus) between the main board (BE-28) and the VIF-40
board.
(e) Error Log Tool
This item is used to check or delete the saved error log.
When the diagnostic item "5. Error Log Tool MENU" in the main menu is selected, the following
submenu appears.
Read Error Log
Reads an error log saved in the EEPROM. When an error log exists, it is judged as an error. In this
case, the cause of the error can be identified by referring to the error details in the error code list.
Delete Error Log
Execute this to delete the error log saved in the EEPROM.
The error log saved in the EEPROM is not deleted automatically. Be sure to delete the error log after
completing the maintenance.
When the deletion starts, the following dialog appears for the final confirmation of deletion.
If the following dialog appears after deleting the error log, the error log is deleted correctly.
1-32 (E)
BCU-100 MM