Managing Climate Control
Designed to be ideal for most patients, Climate Control Auto enables the automatic delivery of a constant
value of absolute humidity while protecting against rainout.
To allow for increased flexibility, Climate Control can be turned to Manual in either the Patient Setup
(when enabled) or the Clinical Setup menus. Setting Climate Control to Manual disables the automatic
control of humidity and allows the patient to set humidity and temperature levels independently.
However, rainout protection is not provided when Climate Control is set to manual.
Mode
Climate Control – Auto
Climate Control –
Manual
* When the temperature setting is set to Off the tube will not heat the air, nor will the humidifier heat the water to
add humidity to the air.
Humidity
Setting range
Default settings
Constant
absolute humidity
(depending on
temperature
setting)
Humidity level
Off–6.0
(0.5 increments)
Setting range
Climate Control
–
Off*, 60ºF–86ºF
(16ºC–30ºC)
3
Off, 60ºF–86ºF
(16ºC–30ºC)
Temperature
Default settings
80ºF (27ºC)
Heated tube
80ºF (27ºC)
25
English