AutoSet Response

AutoSet mode (AirSense 10 AutoSet Plus device only).
For patients who are sensitive to faster changes in pressure during therapy, AutoSet Response can
be set to either Standard or Soft. If set to soft, patients will receive gentler pressure rises during
therapy.
Patients who use the AutoSet Response feature will still get the benefits of ResMed's AutoSet
technology including improved sensitivity to flow-limitation and CSA Detection with Forced
Oscillation Technique.

Climate Control

Climate Control is an intelligent system that controls the humidifier and the ClimateLineAir heated
air tubing to deliver constant, comfortable temperature and humidity levels during therapy.
Designed to prevent dryness of the nose and mouth, it maintains the set temperature and relative
humidity while you sleep. Climate Control can be set to either Auto or Manual and is only available
when both the ClimateLineAir and the HumidAir humidifier are attached.

Climate Control Auto

Climate Control Auto is the recommended and default setting. Climate Control Auto is designed to
make therapy as easy as possible, so there is no need to change the temperature or humidity
settings.
Climate Control adjusts the humidifier output to maintain a constant, comfortable humidity level of
85% relative humidity while protecting against rainout (water droplets in the air tubing and mask).

Tube Temperature

In Climate Control Auto there is no need to change any settings, but if the air in the mask is too
warm or cold for the patient, the tube temperature can be adjusted. The Tube Temperature can be
set to anywhere between 16–30°C, or turned off completely.

Climate Control Manual

Designed to offer more flexibility and control over settings, Climate Control Manual lets the patient
adjust the temperature and humidity to the setting which is most comfortable for them.
In Climate Control Manual, the Tube Temperature and the Humidity Level can be set independently
however, rainout protection is not guaranteed. If rainout does occur, first try increasing the tube
temperature. If the air temperature becomes too warm and rainout continues, try decreasing the
humidity.
Tube Temperature
If the air in the mask feels too warm or too cold, the patient can adjust the temperature to find what
is most comfortable or turn it off completely. The Tube Temperature can be set to anywhere
between16–30°C.
The temperature sensor located at the mask end of the ClimateLineAir heated air tubing enables the
system to automatically control the temperature of the air delivered to the patient. This ensures the
temperature of the air delivered to the patient does not fall below the set minimum temperature,
therefore maximizing breathing comfort for the patient.
English
11