Reset The Motors With Hb75 - Invacare SB 910 User Manual

Hospital bed
Table of Contents

Advertisement

Before use
After 3 months
6
7
X
8
X
9
X
10
X
11
X
12
X
13
X
14
X
15
16
17

6.4 Reset the motors with HB75

The motors need to be reset when:
the hi/lo motor has been replaced
the mattress support is leaning (cannot be set
to horizontal)
Before resetting
1. Disconnect the bed from the mains socket and leave it
without power for at least 10 seconds.
2. Connect all cables to the control unit.
3. Connect the bed to the mains socket.
1568774-H
Year 1–8
Every 6
months
X
X
X
X
X
X
X
Year 1–8
Year 1–8
Every 12
Every 24
months
months
X
X
X
X
X
X
X
X
X
X
Reset
Reset by pressing both sides A of the button, until a
beeping sound is heard (= reset).
After reset - test functions
After resetting; all actuators need to run to end positions:
Press every button, one at a time:
1. Up until it stops
2. Down until it stops
Maintenance
Year 9
Year 10
Every 12
Every 12
months
months
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
19