Cleaning; Safety Precautions When Cleaning; Cleaning By Users - Noraxon Ultium 808 User Manual

2d goniometer smartlead biomechanics research system
Table of Contents

Advertisement

2D Goniometer SmartLead User Manual

8 Cleaning

8.1 Safety Precautions When Cleaning

WARNING
Only use a damp cloth with mild soap and water or isopropyl alcohol to clean the
bottom of the Ultium Sensors.
Do not immerse Ultium Sensors in any water or liquid.

8.2 Cleaning by Users

It is advisable to clean the 2D Goniometer SmartLead on a regular basis. The 2D Goniometer
SmartLead can be cleaned with a cloth slightly dampened with a solution of mild soap and water or
disinfectant solution(i.e. isopropyl alcohol swabs or household disinfectant wipes).
The 2D Goniometer SmartLead is not warranted against exposure to any of the conventional forms of
sterilization(autoclave, heating, etc.). Users wishing to utilize this equipment in a sterile environment,
such as an operating theater, should consult Noraxon for other options.
When cleaning or disinfecting, the sensors must be disconnected from all
instrumentation.
No solvents, acidic or strong alkaline materials should be used to clean the
sensors or damage will result.
To clean the 2D Goniometer, wipe the sensors (green endblocks) with a damp cloth or a cloth
moistened with soapy water.
To disinfect the 2D Goniometer, wipe the sensors (green endblocks) using a cloth moistened with an
allowable disinfectant.
(Rev D January 2020)
19