Technical Speci cations

General
Safe working load (SWL)
Sara Flex
Component Weight
Sara Flex complete
Battery Pack
Electrical
Degree of protection - Electrical Shock
Lift - Protection Class
Hand Control - Protection Class
Battery Type
Battery Part Number
Battery capacity
Battery Charger
Lift - Protection Class - shock protection
Lift nominal voltage
Fuse
Operating force of controls
Max Duty Cycle, Lift Actuator
Max Duty Cycle, Leg Spread Actuator
Maximum sound power level
Scale speci cation
Weight range
Display resolution and type
Operating conditions
Ambient temperature
Relative humidity range
Atmospheric Pressure
Transport and Storage
Ambient temperature
Relative humidity range
Atmospheric Pressure
When transporting the product from one facility to another, make sure the product is properly
secured.
200 kg / 440 lb
52 kg / 114.5 lb
3.8 kg / 8.4 lb
Type BF
IP24
IPX7
Rechargeable - sealed lead acid
NDA0100-20
24V DC 4Ah
NDA8200
Internally powered equipment
24V DC
15A type AGC fuse
< 5 N
10% MAX (2 min ON / 18 min OFF)
10% MAX (2 min ON / 18 min OFF)
50dB
200 kg / 440 lb
0.2 kg / 0.5 lb, liquid crystal display
+10°C (50°F) to +40°C (104°F)
10% to 80%
700 hPa to 1060 hPa
-25°C (-13°F) to +70°C (158°F)
10% to 80%
700 hPa to 1060 hPa
33