Care And Preventive Maintenance - Arjohuntleigh Sara Flex Instructions For Use Manual

Table of Contents

Advertisement

Care and Preventive Maintenance

Sara Flex is subject to wear and tear, and the following actions must be performed when speci ed to
make sure that the product remains within its original manufacturing speci cation.
WARNING
To avoid malfunction resulting in injury, make sure to conduct regular inspections and
follow the recommended maintenance schedule. In some cases, due to heavy use of the
product and exposure to aggressive environment, more frequent inspections should be
carried out.
Local regulations and standards may be more stringent than the recommended
maintenance schedule.
WARNING
To avoid injury to both the patient and the caregiver, never modify the equipment or use
incompatible parts.
SCHEDULE
Caregiver Obligations Action/Check
Cleaning/Disinfection
Examine the sling, straps and clips for damage or fraying as
required.
Check to make sure the patient/resident handles are secure.
Visually check exposed surfaces for damage, sharp edges,
etc.
Visually check sling attachment points. Do not use if
damaged.
Make sure all labels are attached.
Visually check the hand control and cable for damage.
Perform a full functionality test on Sara Flex.
Check operation of the Emergency Stop/Power Button and
the Emergency Lowering Handle
Check battery for leakage and/or deterioration. Replace if
needed.
Check all castors for wear.
Check for evidence of corrosion.
Yearly checks by quali ed service personnel only.
Between
Every Use
29
Every
Week
x
x
x
x
x
x
x
x
x
x
x
x
Every
Year
x

Advertisement

Table of Contents
loading

Table of Contents