Preparation
O
supply via injector
2
1 Press locking lever on mounting flange and push injector
into flange.
Allow locking lever on injector to engage. The injector
is now locked.
Connect up hose.
Fit bed canopy (see page 25).
O
supply should only be used with O
2
monitoring, e.g. via MiniOx 3000.
O
monitor
2
e.g. MiniOx 3000
Prepare the O
monitor in conformity with its specific
2
Instructions for Use.
Fasten O
monitor, complete with holder, to the standard
2
rail. Turn the O
monitor so that the control panel of the
2
measuring unit faces the mattress heater control panel.
Place O
sensor in the warming bed and feed the cable
2
through the hole in the end panel.
Plug the connector of the sensor into the measuring unit.
10
concentration
2
1
Instructions for Use Babytherm 8000 WB