PR 4222 Product Manual page 19

Universal i/f converter
Hide thumbs Also See for 4222:
Table of Contents

Advertisement

Language (LANG): In the menu "LANG" you can choose between 7 different language versions of help texts that will appear
in the menu. You can choose between UK, DE, FR, IT, ES, SE and DK.
Auto diagnosis
The device performs an advanced auto diagnosis of the internal circuits.
The following possible errors can by displayed in the front unit PR 4500.
CJ.ER - CJC sensor defect or CJC temperature outside range
FL.ER - Flash error
NO.CO - Connection error
IN.ER - Error levels on input
TY.ER - Configuration in PR 4500 does not match this product type
Selection of units
After choosing the input signal type you can choose which process units should be displayed in text line 2 (see table). By
selection of temperature input the process value is always displayed in Celsius or Fahrenheit. This is selected in the menu
point after selection of temperature input.