Chapter 4
Analog Input
Table 4-3. MIO X Series Analog Input Configuration
AI Ground-Reference
*
Setting
Differential
Non-Referenced
Single-Ended (NRSE)
Referenced Single-Ended
(RSE)
*
Refer to the
Analog Input Ground-Reference Settings
software considerations.
Refer to the
Connecting Ground-Referenced Signal Sources
X Series User Manual
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Floating Signal Sources
(Not Connected to Building
Ground)
Examples:
• Ungrounded thermocouples
• Signal conditioning with
isolated outputs
• Battery devices
Signal Source
DAQ Device
AI+
+
+
AI–
AI GND
Signal Source
DAQ Device
AI
+
+
AI SENSE
AI GND
Signal Source
DAQ Device
AI
+
+
AI GND
section for descriptions of the RSE, NRSE, and DIFF modes and
section for more information.
4-12
Ground-Referenced
Signal Sources
Example:
• Plug-in instruments with
non-isolated outputs
Signal Source
DAQ Device
AI+
+
+
AI–
AI GND
Signal Source
DAQ Device
AI
+
+
AI SENSE
AI GND
NOT RECOMMENDED
Signal Source
DAQ Device
AI
+
+
V
A
V
B
AI GND
Ground-loop potential (V
– V
) are added
A
B
to measured signal.
ni.com