drives the ADC with this amplified voltage. The amount of amplification
(the gain), is determined by the analog input range, as shown in Figure 4-2.
Table 4-2 shows how signals are routed to the NI-PGIA on MIO X Series
devices.
Table 4-2. Signals Routed to the NI-PGIA on MIO X Series Devices
AI Ground-Reference
Settings
RSE
NRSE
DIFF
For differential measurements, AI 0 and AI 8 are the positive and negative
inputs of differential analog input channel 0. For a complete list of signal
pairs that form differential input channels, refer to the
Descriptions
Caution
The maximum input voltages rating of AI signals with respect to ground (and for
signal pairs in differential mode with respect to each other) are listed in the specifications
document for your device. Exceeding the maximum input voltage of AI signals distorts the
measurement results. Exceeding the maximum input voltage rating also can damage the
device and the computer. NI is not liable for any damage resulting from such signal
connections.
© National Instruments
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
V
in+
PGIA
V
in–
Figure 4-2. MIO X Series Device NI-PGIA
Signals Routed to the Positive
Input of the NI-PGIA (V
in+
AI <0..31>
AI <0..15>
AI <16..31>
AI <0..7>
AI <16..23>
section of Chapter 3,
4-5
Chapter 4
+
Measured
V
m
Voltage
] × Gain
V
= [V
– V
m
in+
in–
Signals Routed to the Negative
)
Input of the NI-PGIA (V
AI GND
AI SENSE
AI SENSE 2
AI <8..15>
AI <24..31>
I/O Connector Signal
Connector and LED
Information.
X Series User Manual
Analog Input
)
in–