National Instruments DAQ X Series User Manual page 211

Hide thumbs Also See for DAQ X Series:
Table of Contents

Advertisement

Chapter 8
PFI
Filter Setting
Filter Clock
None
90 ns
100 MHz
(short)
5.12 μs
100 MHz
(medium)
2.56 ms
100 kHz
(high)
Custom
User
configurable
The filter setting for each input can be configured independently. On power
up, the filters are disabled. Figure 8-3 shows an example of a low to high
transition on an input that has a custom filter set to N = 5.
RTSI, PFI, or
PXI_STAR Terminal
Filter Clock
Filtered Input
Enabling filters introduces jitter on the input signal. The maximum jitter is
one period of the timebase.
When a RTSI input is routed directly to PFI, the X Series device does not
use the filtered version of the input signal.
X Series User Manual
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Table 8-1. Filters
N (Filter Clocks
Needed to
Pass Signal)
9
512
256
N
1
1
2
3
4
1
Figure 8-3. Filter Example
8-6
Pulse Width
Pulse Width
Guaranteed to
Guaranteed to
Pass Filter
Not Pass Filter
90 ns
5.12 μs
2.56 ms
N/timebase
(N – 1)/timebase
Filtered input goes
high when terminal
2
3
4
5
is sampled high on
five consecutive filter
clocks.
80 ns
5.11 μs
2.55 ms