2. Torque Wrench
If you plan to adjust and maintain this chair
yourself, Sunrise recommends that you use
a torque wrench.
Note: A torque setting is the optimum tightening
which should be made on a particular fastener.
A torque wrench which measures in.-lbs. should
be used when tightening fasteners on Breezy
wheelchairs. Unless otherwise noted, use a torque
setting of 50-65 in./lbs. (5.6-7.3 N . m) when setting-
up this wheelchair.
When properly set up, your chair will roll smoothly
and easily. All components should also work
smoothly. If you detect a problem, be sure to
correct it before using this chair.
C. TO FIND & CORRECT A PROBLEM
1. Review the TROUBLESHOOTING section
to see if your chair has one of the listed
symptoms. If so, follow steps to solve the
problem.
2. Follow steps in this section to make sure
set up is correct.
3. If steps 1 and 2 do not work, consult your
authorized supplier for help.
4. If you still have a problem, contact customer
service at 800.456.8165.
D. ANTI-TIP TUBES (Optional)
Anti-tip tubes are recommended unless you are a
skilled rider of this chair and are sure you are not
at risk to tip over.
The following is recommended for all Breezy
models except the 600 (listed next):
Note: When locked in place, anti-tip tube wheels
should be between 11/2" to 2" (38 mm to 51 mm)
off the ground. If set too high, they may not prevent
a tip over. If set too low, they may "hang-up" and
cause a fall or tip over.
930300 Rev. E