Chapter 8 Display System Settings; Waveform Display Setting; Xy Mode - UNI-T UPO3000E Series User Manual

Ultra phosphor oscilloscope
Hide thumbs Also See for UPO3000E Series:
Table of Contents

Advertisement

Chapter 8 Display System Settings

8.1 Waveform Display Setting

You can set the waveform display type, display format, duration, grid brightness,
waveform brightness, and color temperature.
Display Type:
Press DISPLAY
type to select vector or point display.
(1) Vector: This mode provides the most realistic waveforms in most cases,
allowing users to easily view the steep edges of the waveforms (such as
square waves).
(2) Point: Displays the sampling points directly.
Display Format:
Press DISPLAY
format to select YT, XY 1&2, XY 3&4.
(1) YT: Displays voltage value on time scale (horizontal scale).
(2) XY 1&2: Displays the Lissajous figure of CH1~CH2 waveforms, you can
easily measure the phase difference between the two signals of the same
frequency.
(3) XY 3&4: Displays the Lissajous figure of CH3~CH4 waveforms, you can
easily measure the phase difference between the two signals of the same
frequency. (Only for 4-channel models).
Grid Brightness:
Press DISPLAY
grid brightness, and adjust the Multipurpose knob or
shuttle knob to set the grid brightness.
Waveform Brightness:
Press DISPLAY
waveform brightness, and adjust the Multipurpose knob
or shuttle knob to set the waveform brightness.
Persist:
Press DISPLAY
persist to select MIN, 50ms, 100ms, 200ms, 500ms,
1s, 2s, 5s, 10s, 20s or infinite.
Color Temperature:
Press DISPLAY
PgDn
color temperature, select on and it will
directly reflect the probability of waveform signal occurrence. A waveform
with a high frequency of occurrence is displayed in warm colors, and a
waveform with a low frequency of occurrence is displayed in cool color.
Inverse Color Temperature:
Press DISPLAY
PgDn
inverse color temperature and select on,
this function is opposite to the color temperature function.
Note: The inverse color temperature is available only when the color
temperature is set to on.

8.2 XY Mode

The waveform displayed in XY mode is also called the Lissajous figure.
When XY 1&2 are selected, the CH1 signal is input on the horizontal axis
(X axis), and the CH2 signal is input on the vertical axis (Y axis).
When XY 3&4 are selected, the CH3 signal is input on the horizontal axis
(X axis), and the CH4 signal is input on the vertical axis (Y axis). Only
4-channel models have this function.
In X-Y mode, when CH1 or CH3 is activated, use the horizontal POSITION
knob to move the XY figure in the horizontal direction. When CH2 or CH4
is activated, use the vertical POSITION knob to move the XY figure in the
vertical direction.