Summary of Contents for SICK DUSTHUNTER SP100 Ex-3K
Page 1
MMMOPERATING INSTRUCTIONS MMMI O P E R A T I N G I N S T R U C T I O N S DUSTHUNTER SP100 Ex-3K Scattered Light Dust Measuring Device...
Page 2
Original document This document is an original document of SICK Engineering GmbH. O P E R A T I N G I N S T R U C T I O N S | DUSTHUNTER SP100 Ex-3K 8020216/19YN/V3-0/2020-12| SICK Subject to change without notice...
Standard interfaces of the MCU control unit.......38 3.6.3 SOPAS ET user interface ..............38 8020216/19YN/V3-0/2020-12 | SICK O P E R A T I N G I N S T R U C T I O N S | DUSTHUNTER SP100 Ex-3K Subject to change without notice...
Page 4
Connection to the MCUDH Ex-3K control unit......61 Communication options ................62 O P E R A T I N G I N S T R U C T I O N S | DUSTHUNTER SP100 Ex-3K 8020216/19YN/V3-0/2020-12 | SICK...
Page 5
10.5 MCU control unit displays and operating elements........90 8020216/19YN/V3-0/2020-12 | SICK O P E R A T I N G I N S T R U C T I O N S | DUSTHUNTER SP100 Ex-3K Subject to change without notice...
Page 6
Replacing the button cell in the control unit......111 12.8.2 Replacing the Interface module ..........111 O P E R A T I N G I N S T R U C T I O N S | DUSTHUNTER SP100 Ex-3K 8020216/19YN/V3-0/2020-12 | SICK Subject to change without notice...
Page 7
Spare parts, DUSTHUNTER SP100 Ex-3K......... 134 16.2.2 Spare parts, control unit............134 8020216/19YN/V3-0/2020-12 | SICK O P E R A T I N G I N S T R U C T I O N S | DUSTHUNTER SP100 Ex-3K Subject to change without notice...
Page 8
18.2 Electrical protection..................137 18.3 Approvals ....................137 18.4 Licenses ......................138 O P E R A T I N G I N S T R U C T I O N S | DUSTHUNTER SP100 Ex-3K 8020216/19YN/V3-0/2020-12 | SICK Subject to change without notice...
Observe the delivered Product CD as well as other delivered documents. Further information on the internet under: sick.com 8020216/19YN/V3-0/2020-12| SICK O P E R A T I N G I N S T R U C T I O N S | DUSTHUNTER SP100 Ex-3K Subject to change without notice...
Hazard which could result in property damage. Note Hints. O P E R A T I N G I N S T R U C T I O N S | DUSTHUNTER SP100 Ex-3K 8020216/19YN/V3-0/2020-12| SICK Subject to change without notice...
▸ Remove solid and liquid residues. O P E R A T I N G I N S T R U C T I O N S | DUSTHUNTER SP100 Ex-3K 8020216/19YN/V3-0/2020-12| SICK Subject to change without notice...
Page 13
(see “Return delivery”, page 123). 8020216/19YN/V3-0/2020-12| SICK O P E R A T I N G I N S T R U C T I O N S | DUSTHUNTER SP100 Ex-3K Subject to change without notice...
Page 14
124), safe operation of the measuring system is not possible when this value is exceeded. O P E R A T I N G I N S T R U C T I O N S | DUSTHUNTER SP100 Ex-3K 8020216/19YN/V3-0/2020-12| SICK...
Page 15
SICK Service. 8020216/19YN/V3-0/2020-12| SICK O P E R A T I N G I N S T R U C T I O N S | DUSTHUNTER SP100 Ex-3K Subject to change without notice...
Warning: Laser class 2, do not look into the beam. Manufacturer's identification O P E R A T I N G I N S T R U C T I O N S | DUSTHUNTER SP100 Ex-3K 8020216/19YN/V3-0/2020-12| SICK Subject to change without notice...
Information on the specifications of the relay contacts. 8020216/19YN/V3-0/2020-12| SICK O P E R A T I N G I N S T R U C T I O N S | DUSTHUNTER SP100 Ex-3K Subject to change without notice...
– Ignition protection types and applicable standards – Installation regulations and zone classification O P E R A T I N G I N S T R U C T I O N S | DUSTHUNTER SP100 Ex-3K 8020216/19YN/V3-0/2020-12| SICK...
The specified maintenance and inspection tasks are carried out regularly by qualified, experienced personnel. 8020216/19YN/V3-0/2020-12| SICK O P E R A T I N G I N S T R U C T I O N S | DUSTHUNTER SP100 Ex-3K Subject to change without notice...
Page 20
Available for reference. ▸ Passed on to new owners. O P E R A T I N G I N S T R U C T I O N S | DUSTHUNTER SP100 Ex-3K 8020216/19YN/V3-0/2020-12| SICK Subject to change without notice...
Fig. 5: Type plate,MCU control unit (example) 8020216/19YN/V3-0/2020-12| SICK O P E R A T I N G I N S T R U C T I O N S | DUSTHUNTER SP100 Ex-3K Subject to change without notice...
(for further information, see “Selecting the control unit”, page 41). O P E R A T I N G I N S T R U C T I O N S | DUSTHUNTER SP100 Ex-3K 8020216/19YN/V3-0/2020-12| SICK Subject to change without notice...
= 0, cc1 = 1, cc0 = 0). 8020216/19YN/V3-0/2020-12| SICK O P E R A T I N G I N S T R U C T I O N S | DUSTHUNTER SP100 Ex-3K Subject to change without notice...
In addition, ignition protection types “nA” and “nC” are used on the versions with integrated power supply unit. O P E R A T I N G I N S T R U C T I O N S | DUSTHUNTER SP100 Ex-3K 8020216/19YN/V3-0/2020-12| SICK...
8020216/19YN/V3-0/2020-12| SICK O P E R A T I N G I N S T R U C T I O N S | DUSTHUNTER SP100 Ex-3K Subject to change without notice...
Page 26
Ex-marking gas and dust zone 2/22 Maximum surface temperature - T6: 85 °C O P E R A T I N G I N S T R U C T I O N S | DUSTHUNTER SP100 Ex-3K 8020216/19YN/V3-0/2020-12| SICK Subject to change without notice...
(Input, voltage supply 90…250 V) Fig. 9: MCUDH Ex-3K control unit 8020216/19YN/V3-0/2020-12| SICK O P E R A T I N G I N S T R U C T I O N S | DUSTHUNTER SP100 Ex-3K Subject to change without notice...
Page 28
- O: ATEX zone 2/22: Software - E: Emission O P E R A T I N G I N S T R U C T I O N S | DUSTHUNTER SP100 Ex-3K 8020216/19YN/V3-0/2020-12| SICK Subject to change without notice...
Connection terminal, voltage supply input Fig. 10: MCU control unit 8020216/19YN/V3-0/2020-12| SICK O P E R A T I N G I N S T R U C T I O N S | DUSTHUNTER SP100 Ex-3K Subject to change without notice...
Page 30
EX certification - N: None Software - E: Emission O P E R A T I N G I N S T R U C T I O N S | DUSTHUNTER SP100 Ex-3K 8020216/19YN/V3-0/2020-12| SICK Subject to change without notice...
135). The weather hood housing available is suitable for use in the Ex-zone. Fig. 12: Weather hood housing for the Ex-area O P E R A T I N G I N S T R U C T I O N S | DUSTHUNTER SP100 Ex-3K 8020216/19YN/V3-0/2020-12| SICK...
(e.g., due to an electronic defect). An error message is generated when the “zero point” is outside the specified range. 8020216/19YN/V3-0/2020-12| SICK O P E R A T I N G I N S T R U C T I O N S | DUSTHUNTER SP100 Ex-3K Subject to change without notice...
Page 34
Receiver optics in control position Fig. 14: Contamination and control value measurement O P E R A T I N G I N S T R U C T I O N S | DUSTHUNTER SP100 Ex-3K 8020216/19YN/V3-0/2020-12| SICK...
Dust deposits must be removed regularly and properly. 8020216/19YN/V3-0/2020-12| SICK O P E R A T I N G I N S T R U C T I O N S | DUSTHUNTER SP100 Ex-3K Subject to change without notice...
MCUDH Ex-3K (control unit for zone 2/22) Fig. 15: Usage illustration with application variants DUSTHUNTER SP100 Ex-3K O P E R A T I N G I N S T R U C T I O N S | DUSTHUNTER SP100 Ex-3K 8020216/19YN/V3-0/2020-12| SICK...
SOPAS ET is delivered on the product CD. Alternatively, you can download SOPAS ET free of charge from the SICK homepage (“Downloads”). O P E R A T I N G I N S T R U C T I O N S | DUSTHUNTER SP100 Ex-3K 8020216/19YN/V3-0/2020-12| SICK...
– NOx: 1000 mg/m³ – HF: 10 mg/m³ 8020216/19YN/V3-0/2020-12| SICK O P E R A T I N G I N S T R U C T I O N S | DUSTHUNTER SP100 Ex-3K Subject to change without notice...
Fixing bolts Fig. 17: Flange with tube (4 bolts) O P E R A T I N G I N S T R U C T I O N S | DUSTHUNTER SP100 Ex-3K 8020216/19YN/V3-0/2020-12| SICK Subject to change without notice...
Further MCU control unit variants available on request 8020216/19YN/V3-0/2020-12| SICK O P E R A T I N G I N S T R U C T I O N S | DUSTHUNTER SP100 Ex-3K Subject to change without notice...
All dimensions in mm Fig. 19: MCUDH Ex-3K control unit clearance 8020216/19YN/V3-0/2020-12| SICK O P E R A T I N G I N S T R U C T I O N S | DUSTHUNTER SP100 Ex-3K Subject to change without notice...
Page 44
Line clearance Fig. 20: MCU control unit clearance O P E R A T I N G I N S T R U C T I O N S | DUSTHUNTER SP100 Ex-3K 8020216/19YN/V3-0/2020-12| SICK Subject to change without notice...
The maintenance cycle can thus be considerably extended. 8020216/19YN/V3-0/2020-12| SICK O P E R A T I N G I N S T R U C T I O N S | DUSTHUNTER SP100 Ex-3K Subject to change without notice...
Data (see “Technical data”, page 124) fuses O P E R A T I N G I N S T R U C T I O N S | DUSTHUNTER SP100 Ex-3K 8020216/19YN/V3-0/2020-12| SICK Subject to change without notice...
(see “Technical data”, page 124). 8020216/19YN/V3-0/2020-12| SICK O P E R A T I N G I N S T R U C T I O N S | DUSTHUNTER SP100 Ex-3K Subject to change without notice...
Scope of delivery Check the scope of delivery according to the order confirmation. O P E R A T I N G I N S T R U C T I O N S | DUSTHUNTER SP100 Ex-3K 8020216/19YN/V3-0/2020-12| SICK...
“Fitting and connecting the sender/receiver unit”, page 66). 8020216/19YN/V3-0/2020-12| SICK O P E R A T I N G I N S T R U C T I O N S | DUSTHUNTER SP100 Ex-3K Subject to change without notice...
(see “Fastening technology”, page 135) to prevent gas escaping. O P E R A T I N G I N S T R U C T I O N S | DUSTHUNTER SP100 Ex-3K 8020216/19YN/V3-0/2020-12| SICK Subject to change without notice...
Shown for high-temperature version, nominal length 735 mm 8020216/19YN/V3-0/2020-12| SICK O P E R A T I N G I N S T R U C T I O N S | DUSTHUNTER SP100 Ex-3K Subject to change without notice...
Base plate Fig. 25: Fitting the weather hood housing for the sender/receiver unit O P E R A T I N G I N S T R U C T I O N S | DUSTHUNTER SP100 Ex-3K 8020216/19YN/V3-0/2020-12| SICK...
For outdoor installation, it is necessary for the customer to provide a weather hood. 8020216/19YN/V3-0/2020-12| SICK O P E R A T I N G I N S T R U C T I O N S | DUSTHUNTER SP100 Ex-3K Subject to change without notice...
During commissioning: Establish the connections still required on the sender/receiver unit at the measuring point. O P E R A T I N G I N S T R U C T I O N S | DUSTHUNTER SP100 Ex-3K 8020216/19YN/V3-0/2020-12| SICK...
Plug Socket +24 V Fig. 26: Connection plan 8020216/19YN/V3-0/2020-12| SICK O P E R A T I N G I N S T R U C T I O N S | DUSTHUNTER SP100 Ex-3K Subject to change without notice...
(see “Fitting and connecting the sender/receiver unit”, page 66). O P E R A T I N G I N S T R U C T I O N S | DUSTHUNTER SP100 Ex-3K 8020216/19YN/V3-0/2020-12| SICK Subject to change without notice...
Fig. 27: MCUDH Ex-3K control unit component layout 8020216/19YN/V3-0/2020-12| SICK O P E R A T I N G I N S T R U C T I O N S | DUSTHUNTER SP100 Ex-3K Subject to change without notice...
Optional Interface module Fig. 28: MCU component layout O P E R A T I N G I N S T R U C T I O N S | DUSTHUNTER SP100 Ex-3K 8020216/19YN/V3-0/2020-12| SICK Subject to change without notice...
Only modify wiring when disconnected from the power supply and potential-free. Fig. 29: MCUDH Ex-3K line glands 8020216/19YN/V3-0/2020-12| SICK O P E R A T I N G I N S T R U C T I O N S | DUSTHUNTER SP100 Ex-3K Subject to change without notice...
Fig. 31: MCU processor board connections O P E R A T I N G I N S T R U C T I O N S | DUSTHUNTER SP100 Ex-3K 8020216/19YN/V3-0/2020-12| SICK Subject to change without notice...
● – Communication: Connection terminals 38 and 39 8020216/19YN/V3-0/2020-12| SICK O P E R A T I N G I N S T R U C T I O N S | DUSTHUNTER SP100 Ex-3K Subject to change without notice...
RS485 Fig. 33: Communication options 1 and 2 O P E R A T I N G I N S T R U C T I O N S | DUSTHUNTER SP100 Ex-3K 8020216/19YN/V3-0/2020-12| SICK Subject to change without notice...
Page 63
Fig. 34: Communication options 3, 4 and 5 8020216/19YN/V3-0/2020-12| SICK O P E R A T I N G I N S T R U C T I O N S | DUSTHUNTER SP100 Ex-3K Subject to change without notice...
Page 64
24 V (6075027) Fig. 35: MCUDH Ex-3K interface options O P E R A T I N G I N S T R U C T I O N S | DUSTHUNTER SP100 Ex-3K 8020216/19YN/V3-0/2020-12| SICK...
Measuring point has been checked for free access without problems or hazards. ● 8020216/19YN/V3-0/2020-12| SICK O P E R A T I N G I N S T R U C T I O N S | DUSTHUNTER SP100 Ex-3K Subject to change without notice...
Ensure the device can be switched off with a power isolating switch or circuit breaker in accordance with EN 61010-1. 8020216/19YN/V3-0/2020-12| SICK O P E R A T I N G I N S T R U C T I O N S | DUSTHUNTER SP100 Ex-3K Subject to change without notice...
Preset password: sickoptic O P E R A T I N G I N S T R U C T I O N S | DUSTHUNTER SP100 Ex-3K 8020216/19YN/V3-0/2020-12| SICK Subject to change without notice...
Enter a name for this search if you want to save it. Click “Finish”. 8020216/19YN/V3-0/2020-12| SICK O P E R A T I N G I N S T R U C T I O N S | DUSTHUNTER SP100 Ex-3K Subject to change without notice...
Enter a name for this search if you want to save it. Click “Finish”. O P E R A T I N G I N S T R U C T I O N S | DUSTHUNTER SP100 Ex-3K 8020216/19YN/V3-0/2020-12| SICK...
Configure the Interface module according to requirements. Fig. 37: SOPAS ET menu: MCU/Configuration/System Configuration 8020216/19YN/V3-0/2020-12| SICK O P E R A T I N G I N S T R U C T I O N S | DUSTHUNTER SP100 Ex-3K Subject to change without notice...
Assign a name for this search. Click “Finish”. O P E R A T I N G I N S T R U C T I O N S | DUSTHUNTER SP100 Ex-3K 8020216/19YN/V3-0/2020-12| SICK Subject to change without notice...
GSD file and measured value assignment are available for the PROFIBUS DP Interface module on request. 8020216/19YN/V3-0/2020-12| SICK O P E R A T I N G I N S T R U C T I O N S | DUSTHUNTER SP100 Ex-3K Subject to change without notice...
“Expansion Module Information” field. Fig. 39: SOPAS ET menu: MCU/Configuration/System Configuration O P E R A T I N G I N S T R U C T I O N S | DUSTHUNTER SP100 Ex-3K 8020216/19YN/V3-0/2020-12| SICK Subject to change without notice...
Fig. 41: SOPAS ET menu: DH SP100/Configuration/Application parameters (top window) 8020216/19YN/V3-0/2020-12| SICK O P E R A T I N G I N S T R U C T I O N S | DUSTHUNTER SP100 Ex-3K Subject to change without notice...
Set the sender/receiver unit to “Maintenance”: Click “Maintenance sensor”. Fig. 43: SOPAS ET menu: MCU/Maintenance/Maintenance O P E R A T I N G I N S T R U C T I O N S | DUSTHUNTER SP100 Ex-3K 8020216/19YN/V3-0/2020-12| SICK...
1 password set and the “Maintenance” mode set. 8020216/19YN/V3-0/2020-12| SICK O P E R A T I N G I N S T R U C T I O N S | DUSTHUNTER SP100 Ex-3K Subject to change without notice...
The value measured last is output during control value determination (see “Function check”, page 33). O P E R A T I N G I N S T R U C T I O N S | DUSTHUNTER SP100 Ex-3K 8020216/19YN/V3-0/2020-12| SICK Subject to change without notice...
Fig. 46: SOPAS ET menu: MCU/Configuration/IO configuration/Output Parameters 8020216/19YN/V3-0/2020-12| SICK O P E R A T I N G I N S T R U C T I O N S | DUSTHUNTER SP100 Ex-3K Subject to change without notice...
Page 80
Value Defines a tolerance for resetting the limit value relay. O P E R A T I N G I N S T R U C T I O N S | DUSTHUNTER SP100 Ex-3K 8020216/19YN/V3-0/2020-12| SICK Subject to change without notice...
This parameter opens the “Analog input 4 - Oxygen” field to set the lower and upper range limit values and the Live Zero value. 8020216/19YN/V3-0/2020-12| SICK O P E R A T I N G I N S T R U C T I O N S | DUSTHUNTER SP100 Ex-3K Subject to change without notice...
Determine regression coefficients from the mA values of the analog output for “Scattered light intensity” and the actual dust concentrations measured gravimetrically. O P E R A T I N G I N S T R U C T I O N S | DUSTHUNTER SP100 Ex-3K 8020216/19YN/V3-0/2020-12| SICK...
Page 83
After entering, set the sender/receiver unit to “Measurement” mode again. 8020216/19YN/V3-0/2020-12| SICK O P E R A T I N G I N S T R U C T I O N S | DUSTHUNTER SP100 Ex-3K Subject to change without notice...
Display settings” directory. Fig. 49: SOPAS ET menu: MCU/Configuration/Display Settings O P E R A T I N G I N S T R U C T I O N S | DUSTHUNTER SP100 Ex-3K 8020216/19YN/V3-0/2020-12| SICK Subject to change without notice...
Page 85
Value 1 - Control unit (MCU) Concentration s.c.dry O2 corr. (SI) 8020216/19YN/V3-0/2020-12| SICK O P E R A T I N G I N S T R U C T I O N S | DUSTHUNTER SP100 Ex-3K Subject to change without notice...
A new COM port is shown. Use this COM port for communication. O P E R A T I N G I N S T R U C T I O N S | DUSTHUNTER SP100 Ex-3K 8020216/19YN/V3-0/2020-12| SICK Subject to change without notice...
SOPAS ET (see “Changing display settings”, page 84). 8020216/19YN/V3-0/2020-12| SICK O P E R A T I N G I N S T R U C T I O N S | DUSTHUNTER SP100 Ex-3K Subject to change without notice...
Status bar Fig. 50: Function elements, LC display, MCUDH Ex-3K control unit O P E R A T I N G I N S T R U C T I O N S | DUSTHUNTER SP100 Ex-3K 8020216/19YN/V3-0/2020-12| SICK...
Sets plus or minus. Confirmation button Confirms the entry. 8020216/19YN/V3-0/2020-12| SICK O P E R A T I N G I N S T R U C T I O N S | DUSTHUNTER SP100 Ex-3K Subject to change without notice...
Current button function Fig. 51: Functional elements LC-Display MCU control unit O P E R A T I N G I N S T R U C T I O N S | DUSTHUNTER SP100 Ex-3K 8020216/19YN/V3-0/2020-12| SICK Subject to change without notice...
Set limit high in °C / hPa (password required) 8020216/19YN/V3-0/2020-12| SICK O P E R A T I N G I N S T R U C T I O N S | DUSTHUNTER SP100 Ex-3K Subject to change without notice...
Display sensor error messages Warnings Display sensor warning messages O P E R A T I N G I N S T R U C T I O N S | DUSTHUNTER SP100 Ex-3K 8020216/19YN/V3-0/2020-12| SICK Subject to change without notice...
82) with the operating fields and save to the device with “Save”. 8020216/19YN/V3-0/2020-12| SICK O P E R A T I N G I N S T R U C T I O N S | DUSTHUNTER SP100 Ex-3K Subject to change without notice...
Example for Parameter protocol Fig. 53: DUSTHUNTER SP100 Parameter protocol (example) O P E R A T I N G I N S T R U C T I O N S | DUSTHUNTER SP100 Ex-3K 8020216/19YN/V3-0/2020-12| SICK Subject to change without notice...
▸ Do not interrupt the purge gas supply. 8020216/19YN/V3-0/2020-12| SICK O P E R A T I N G I N S T R U C T I O N S | DUSTHUNTER SP100 Ex-3K Subject to change without notice...
Cover screw, cleaning opening Fig. 54: Cleaning the optical surfaces O P E R A T I N G I N S T R U C T I O N S | DUSTHUNTER SP100 Ex-3K 8020216/19YN/V3-0/2020-12| SICK Subject to change without notice...
Resume measuring operation (see “Resuming measuring operation”, page 94). 8020216/19YN/V3-0/2020-12| SICK O P E R A T I N G I N S T R U C T I O N S | DUSTHUNTER SP100 Ex-3K Subject to change without notice...
136) can also be used to supply power to the sender/receiver unit without using the control unit. O P E R A T I N G I N S T R U C T I O N S | DUSTHUNTER SP100 Ex-3K 8020216/19YN/V3-0/2020-12| SICK...
“Spare parts, DUSTHUNTER SP100 Ex- 3K”, page 134. 8020216/19YN/V3-0/2020-12| SICK O P E R A T I N G I N S T R U C T I O N S | DUSTHUNTER SP100 Ex-3K Subject to change without notice...
Put the complete device back into operation (switch on voltage and check measured and status values). Fig. 60: Checking laser alignment O P E R A T I N G I N S T R U C T I O N S | DUSTHUNTER SP100 Ex-3K 8020216/19YN/V3-0/2020-12| SICK Subject to change without notice...
Put the complete device back into operation (switch on voltage and check measured and status values). 8020216/19YN/V3-0/2020-12| SICK O P E R A T I N G I N S T R U C T I O N S | DUSTHUNTER SP100 Ex-3K Subject to change without notice...
Put the complete device back into operation (switch on voltage and check measured and status values). O P E R A T I N G I N S T R U C T I O N S | DUSTHUNTER SP100 Ex-3K 8020216/19YN/V3-0/2020-12| SICK...
Put the complete device back into operation (switch on voltage and check measured and status values). 8020216/19YN/V3-0/2020-12| SICK O P E R A T I N G I N S T R U C T I O N S | DUSTHUNTER SP100 Ex-3K Subject to change without notice...
Put the complete device back into operation (switch on voltage and check measured and status values). O P E R A T I N G I N S T R U C T I O N S | DUSTHUNTER SP100 Ex-3K 8020216/19YN/V3-0/2020-12| SICK...
Put the complete device back into operation (switch on voltage and check measured and status values). 8020216/19YN/V3-0/2020-12| SICK O P E R A T I N G I N S T R U C T I O N S | DUSTHUNTER SP100 Ex-3K Subject to change without notice...
MCUDH Ex-3K control unit Fig. 61: Replacing the MCUDH Ex-3K button cell O P E R A T I N G I N S T R U C T I O N S | DUSTHUNTER SP100 Ex-3K 8020216/19YN/V3-0/2020-12| SICK Subject to change without notice...
Fig. 62: Replacing the MCUDH Ex-3K power supply unit 8020216/19YN/V3-0/2020-12| SICK O P E R A T I N G I N S T R U C T I O N S | DUSTHUNTER SP100 Ex-3K Subject to change without notice...
Ribbon cable connector Fig. 63: Replacing the MCUDH Ex-3K Interface module O P E R A T I N G I N S T R U C T I O N S | DUSTHUNTER SP100 Ex-3K 8020216/19YN/V3-0/2020-12| SICK Subject to change without notice...
Ribbon cable connector Fig. 65: Replacing the MCU Interface module 8020216/19YN/V3-0/2020-12| SICK O P E R A T I N G I N S T R U C T I O N S | DUSTHUNTER SP100 Ex-3K Subject to change without notice...
▸ The sender/receiver unit enclosure may only be opened by SICK Service. O P E R A T I N G I N S T R U C T I O N S | DUSTHUNTER SP100 Ex-3K 8020216/19YN/V3-0/2020-12| SICK Subject to change without notice...
“Diagnosis” after selecting the device (e.g. “MCU control unit” or “DH SP100 Ex”). 8020216/19YN/V3-0/2020-12| SICK O P E R A T I N G I N S T R U C T I O N S | DUSTHUNTER SP100 Ex-3K Subject to change without notice...
> ±2%. Contact SICK Service. trol value determination O P E R A T I N G I N S T R U C T I O N S | DUSTHUNTER SP100 Ex-3K 8020216/19YN/V3-0/2020-12| SICK Subject to change without notice...
Page 115
Description of error or warning: In the description field of SOPAS ET. ● 8020216/19YN/V3-0/2020-12| SICK O P E R A T I N G I N S T R U C T I O N S | DUSTHUNTER SP100 Ex-3K Subject to change without notice...
MCU(DH Ex) in Test mode. enabled nance” directory) O P E R A T I N G I N S T R U C T I O N S | DUSTHUNTER SP100 Ex-3K 8020216/19YN/V3-0/2020-12| SICK Subject to change without notice...
Page 117
Description of error or warning: In the description field of SOPAS ET. ● 8020216/19YN/V3-0/2020-12| SICK O P E R A T I N G I N S T R U C T I O N S | DUSTHUNTER SP100 Ex-3K Subject to change without notice...
Put the complete device back into operation (switch on voltage and check measured and status values). O P E R A T I N G I N S T R U C T I O N S | DUSTHUNTER SP100 Ex-3K 8020216/19YN/V3-0/2020-12| SICK...
Fastening screws Fig. 68: Checking the beam path 8020216/19YN/V3-0/2020-12| SICK O P E R A T I N G I N S T R U C T I O N S | DUSTHUNTER SP100 Ex-3K Subject to change without notice...
Fig. 69: Replacing the MCUDH Ex fuse (shown with fuse cover removed) O P E R A T I N G I N S T R U C T I O N S | DUSTHUNTER SP100 Ex-3K 8020216/19YN/V3-0/2020-12| SICK...
Information on sending a device to the factory for inspection or repair, see “Return deliv- ery”, page 123. 8020216/19YN/V3-0/2020-12| SICK O P E R A T I N G I N S T R U C T I O N S | DUSTHUNTER SP100 Ex-3K Subject to change without notice...
Disconnect the control unit from the supply voltage. ▸ Disconnect the potential equalization. O P E R A T I N G I N S T R U C T I O N S | DUSTHUNTER SP100 Ex-3K 8020216/19YN/V3-0/2020-12| SICK Subject to change without notice...
All parts in contact with the sample gas can be contaminated with harmful ● substances. 8020216/19YN/V3-0/2020-12| SICK O P E R A T I N G I N S T R U C T I O N S | DUSTHUNTER SP100 Ex-3K Subject to change without notice...
Contamination limits: At 30% warning, at 40% malfunction Manual linearity test via reference filter O P E R A T I N G I N S T R U C T I O N S | DUSTHUNTER SP100 Ex-3K 8020216/19YN/V3-0/2020-12| SICK...
Page 125
Lappkabel, Unitronic Li2YCY v (TP) 2 mm²×2 mm²×0.5 mm² Burning behavior Flame-retardant acc. to IEC 60332-1-2 8020216/19YN/V3-0/2020-12| SICK O P E R A T I N G I N S T R U C T I O N S | DUSTHUNTER SP100 Ex-3K Subject to change without notice...
Page 126
Version with power supply unit MCU NWxxx I O P E R A T I N G I N S T R U C T I O N S | DUSTHUNTER SP100 Ex-3K 8020216/19YN/V3-0/2020-12| SICK Subject to change without notice...
Page 127
Version without power supply unit MCUDH Ex-3K N2xxx 8020216/19YN/V3-0/2020-12| SICK O P E R A T I N G I N S T R U C T I O N S | DUSTHUNTER SP100 Ex-3K Subject to change without notice...
Sender/receiver unit DHSP-T4V21EX3KT6 1091013 Sender/receiver unit DHSP-T2V21EX3KT6 1091011 O P E R A T I N G I N S T R U C T I O N S | DUSTHUNTER SP100 Ex-3K 8020216/19YN/V3-0/2020-12| SICK Subject to change without notice...
Flange with tube, Øi 70 mm, L 500 mm, with 3.1 Material Certificate 2115418 DHSP-Txx2 8020216/19YN/V3-0/2020-12| SICK O P E R A T I N G I N S T R U C T I O N S | DUSTHUNTER SP100 Ex-3K Subject to change without notice...
24 V DC, without integrated purge gas supply, with display, without integrated Interface module O P E R A T I N G I N S T R U C T I O N S | DUSTHUNTER SP100 Ex-3K 8020216/19YN/V3-0/2020-12| SICK...
MCU control unit Fig. 75: Dimensions, MCU-N control unit 8020216/19YN/V3-0/2020-12| SICK O P E R A T I N G I N S T R U C T I O N S | DUSTHUNTER SP100 Ex-3K Subject to change without notice...
Page 132
- 1 PROFIBUS, RS485, PULS Interface module - 1 Ethernet, CoLa-B, Service Interface module O P E R A T I N G I N S T R U C T I O N S | DUSTHUNTER SP100 Ex-3K 8020216/19YN/V3-0/2020-12| SICK...
For the MCUDH Ex-3K, the use of a button cell other than the one marked as associated in the Table is not permitted. 8020216/19YN/V3-0/2020-12| SICK O P E R A T I N G I N S T R U C T I O N S | DUSTHUNTER SP100 Ex-3K Subject to change without notice...
Fuse sets, T 2 A 250 V (suitable for MCU control unit) 2054541 O P E R A T I N G I N S T R U C T I O N S | DUSTHUNTER SP100 Ex-3K 8020216/19YN/V3-0/2020-12| SICK...
Blind flange for temporary closure of the flange with tube 8020216/19YN/V3-0/2020-12| SICK O P E R A T I N G I N S T R U C T I O N S | DUSTHUNTER SP100 Ex-3K Subject to change without notice...
2049045 DHSP control filter set for linearity check O P E R A T I N G I N S T R U C T I O N S | DUSTHUNTER SP100 Ex-3K 8020216/19YN/V3-0/2020-12| SICK Subject to change without notice...
Please enter as subject “Open Source Soft- ware”. O P E R A T I N G I N S T R U C T I O N S | DUSTHUNTER SP100 Ex-3K 8020216/19YN/V3-0/2020-12| SICK...
Page 139
ANNEX 8020216/19YN/V3-0/2020-12| SICK O P E R A T I N G I N S T R U C T I O N S | DUSTHUNTER SP100 Ex-3K Subject to change without notice...
Page 140
Phone +7 495 283 09 90 E-Mail info@sick.com E-Mail office@sick.com.gr E-Mail info@sick.ru Vietnam Hong Kong Singapore Phone +65 6744 3732 Phone +852 2153 6300 Phone +65 6744 3732 E-Mail sales.gsg@sick.com E-Mail ghk@sick.com.hk E-Mail sales.gsg@sick.com Detailed addresses and further locations at www.sick.com...
Need help?
Do you have a question about the DUSTHUNTER SP100 Ex-3K and is the answer not in the manual?
Questions and answers