Restore Factory Settings - PCE Health and Fitness PCE-TUM 50 Instruction Manual

Turbidity meter
Table of Contents

Advertisement

Unlock or reset the password
6.1
Press and hold the
key for 3 seconds, the meter shows the "Password Protection" and waits for entering the correct digits.
6.2 Press the ◄ or ► key to enter the password, press the
option and "Disable". The password will be removed.

Restore factory settings

The Factory Reset is used to restore the meter back to factory default settings. If enabled, all of the calibration data and selected options/parameters will be
lost or reset, the meter must be recalibrated.
7.1
Press and hold the
key for 3 seconds to enter the setup menu.
Press the ◄ or ► key to select the "Reset", press the
7.2
Press the ◄ or ► key to select "Restore factory settings", press the
7.3
Enter
7.4
Press the
key, the meter immediately restores the factory settings, press the
During the setting process, press the
Enter
key to confirm. Once you have successfully entered the setup menu, selecting the "Password"
Enter
key to confirm.
ESC
key, the meter will exit the setting and return to the measurement mode.
Enter
key. The meter shows "Are you sure you want to restore factory settings ?".
ESC
key to cancel.