Checking The Grayscale Check Results; Setting Whether To Display / Hide Warnings Or Qc History - Eizo RadiForce MX216 Installation Manual

Color lcd monitor
Hide thumbs Also See for RadiForce MX216:
Table of Contents

Advertisement

Checking the Grayscale Check results

Attention
• The results can be checked when "QC History" is set to "On".
Procedure
1. Select "RadiCS SelfQC" from the Setting menu, then select
2. Select "Grayscale Check", then
3. Select "Result", then
4. Select "DICOM", "CAL1", or "CAL2", then select
Up to five past Grayscale Check results (Passed, Failed, Canceled, or Error) are displayed.
3-3. Setting Whether to Display / Hide Warnings or QC
History
You can set whether to display or hide warnings related to "Failed" Grayscale Check results, or the
execution results (QC History) of SelfCalibration, Grayscale Check, etc.
Procedure
1. Select "RadiCS SelfQC" from the Setting menu, then select
2. Select "Settings", then
3. Select "Warning" or "QC History".
4. Select "On" or "Off", then select
.
.
.
.
.
.
.
Chapter 3 Controlling Monitor Quality