Normal start
If the dryer has not been used for more than 3 months, refer to section Initial startup. In all other
cases, proceed as follows:
If the dryer has not been used for more than 3 months, refer to section Initial start-up. In all other
cases, proceed as follows:
1. Cut off the air supply from the compressor towards the dryer by closing the inlet valves.
2. Close the outlet valves between the dryer and the air consumer.
3. Start the compressors and slowly open the inlet valve.
4. Ensure the dryer is powered on.
5. Gradually open the air outlet valves.
4233 5000 13
Close the external inlet valve in case the compressor needs to be restarted. The high air
speed in the startup phase of the compressor may damage the desiccant.
All valves, couplings and pipes used/installed after the unit must be free from oil and
grease (hydrocarbon contamination below 550 mg/m
50 micrometer).
For the location of the external inlet valve and external outlet valve (customer's
installation), see section
Installation
proposal.
2
) and dust (particles size below
61