Reference figure above and the Purge Plug Identification Table on page 78.
Once the correct purge body (1, 2 or 3 hole) and orifice disc (01, 02, 03 or 04) has been selected
per the instructions on Page 78, the purge valve may be assembled.
The index letters located on the purge valve body correspond with the selected orifice sizes per
the Table on Page 78.
Place the appropriate orifice disc onto the valve body, taking care to align the notches. The disc
will only fit in one position. Place the valve stem O-ring(s) into their corresponding recesses on
the back of the valve stem.
Align the holes in the purge sealing gasket with the corresponding holes in the valve stem.
Rotate the valve stem assembly so that the correct orifice letters align with the notches
corresponding to the holes (1, 2 or 3) in the valve stem.
Press the assembly together and fasten with the M4 screw.
Double check to make sure that the open orifice holes correspond with the correct orifice
selection as identified earlier.