CAN settings
For proper working the CAN settings need to be defined. To verify, the following steps can be
followed after entering the Settings submenu and then selecting Network and CAN:
The CAN address should be set to 30. The CAN settings of the compressors and dryers are set
on their controller (see DIN rail controller).
Communication profile should be set to Mk4.
When selecting Ethernet in the Network submenu (see picture above) it is possible to set a
custom IP address, gateway and subnet mask. After connecting an ethernet cable between your
network and the controller, the plant can be monitored on-line when browsing to the set IP
address.
Setting the unit of pressure
In the Settings - General submenu (full access needed, see above), navigate to <Pressure Unit>
and select the desired pressure unit.
4233 5000 13
29