Purge Plug Dmed 025 & 035 - BeaconMedaes HTM Instruction Book

Dual medical air purifiers
Table of Contents

Advertisement

8.6
Purge plug identification/changeout dMED 25 & 35
Ensure shutdown and start up procedures are followed prior to carrying out any purge
plug changeout
Purge Plug Identification
Model
Operating Pressure (barg)
4
5
dMED025
PPK020
PPK018
dMED035
PPK028
PPK023
74
6
7
8
PPK016
PPK015
PPK014
PPK021
PPK020
PPK018
Remove dryer front panel.
Remove purge plug securing screw from the upper valve block.
Remove purge plug.
Replace with appropriate purge plug according to inlet pressure.
Refer to table above.
Grease purge plug O-ring to ensure ease of fit
9
10
11
PPK014
PPK013
PPK013
PPK013
PPK018
PPK016
PPK015
PPK015
1. Securing screw
2. Purge plug
12
13
14
PPK012
PPK012
PPK012
PPK014
PPK014
PPK014
15
16
PPK012
PPK014
4233 5000 13