Reference figure 4.13 and the Purge Plug Identification Table on Page 20.
Once the correct purge body (1, 2 or 3 hole) and orifice disc (01, 02, 03 or 04) has been
selected per the instructions on Page 20, the purge valve may be assembled.
The index letters located on the purge valve body correspond with the selected orifice sizes
per the Table on Page 20.
Place the appropriate orifice disc onto the valve body, taking care to align the notches. The
disc will only fit in one position.
Place the valve stem O-ring(s) into their corresponding recesses on the back of the valve
stem.
Align the holes in the purge sealing gasket with the corresponding holes in the valve stem.
Rotate the valve stem assembly so that the correct orifice letters align with the notches
corresponding to the holes (1, 2 or 3) in the valve stem.
Press the assembly together and fasten with the M4 screw.
Double check to make sure that the open orifice holes correspond with the correct orifice
selection as identified earlier
56
4233 5000 13