Operating Instructions - BeaconMedaes HTM Instruction Book

Dual medical air purifiers
Table of Contents

Advertisement

6

Operating instructions

Safety
Startup valve positions
10
3
5
1
Inlet Valve (Closed)
3
DP sensor valve (Closed)
5
Drain Valve (Closed)
7
Sampling point Valve (Closed)
9
Line Valve (Closed)
Note – Item 3 (PD sensor valve) is for diagnostics only. When closed the dew point reading is taken from line. By
opening this valve, you can read the dewpoint between the dryer column and QDT filter. This allow you to determine if
a dew point fault is due to the dryer column not performing or the QDT has become saturated.
4233 5000 13
Always observe all relevant safety instructions.
4
1
2
Outlet Valve (Open)
4
2 x Link valve (Open)
6
Full flow test point valve (Closed)
8
Sensor maintenance valve (Open)
10
PD flow control valve (Leave as set)
9
6
7
8
2
59