V. 2.0 04/2020 Instruction manual spring testing system Congratulations on your purchase of the SAUTER TVO or the THM test bench. We hope you enjoy your quality measurement system with its wide range of functions and high reproducibility. If operated correctly, this high-quality product will give you many years of use.
(up to 1000 cycles). It can be used for material testing in the metal, plastics and textile industries. It can also be operated with SAUTER software (AFH) and can be conveniently controlled from there using a PC. This software is also able to document force, time and distance.
Page 4
(Possible destruction of the force gauge) • Under no circumstances should you attempt to open, repair or modify the machine. Please contact SAUTER GmbH. • The test stand is not suitable for operation in a humid environment. Avoid penetration of moisture into the housing under all circumstances.
4 Technical data Test bench 50N100 100N100 200N100 300N100 500N100 Maximum force 50 N 100 N 200 N 300 N 500N Readability 0,01N 0,02N 0,05N 0,1N 0,1N Maximum stroke 100mm length Maximum measuring room 100mm height Working 20±10°C temperature Storage and -5°C~40°C transport temp.
5.2 Control keys ON / OFF: On / Off button (press the key for approx. 1 second) UNIT: Measuring units - Press the key briefly: Selection between N, kg and lb ZERO: Zero position Assignment with three functions: - Zero setting of the display (tare function) - Zeroing of the peak value (Peak) - Saving a setting (in SET mode) SET:...
PEAK: (peak value) Assignment with three functions: - continuous measurement - Peak (peak value recording) - Auto-Peak, like Peak function, but without minimum limit function (see section 2.6) MEMORY: (memory function) Saves the peak values for calculating the average of the measurement results (see section - Saving) DELETE FUNCTION: of memory values (only in "Memory"...
5.6 Auto Peak Hold Mode (Auto Peak Mode) Switch over by: 5.7 Minimum limit value function to activate the measured value storage This function is used for measurements where unwanted "pre-peaks" occur which are below the peak value (Fp) being searched for. The adjustable limit value (Fo) prevents the measuring instrument from storing the "pre-peaks".
It is used to restart the unit after an operating error. Memory values and settings are deleted. 6 Thread of the fine adjustment screw This applies to the following SD's: SD 50, 100, 200, 300, 500 The thread of the fine adjustment screw is a DIN M6 thread, according to this DIN is also the pitch.
The device may only be used in accordance with the described specifications. SAUTER must give written approval of any different areas of use / application. Warranty Warranty expires with:...
Page 11
(www.KERN-sohn.com). Observe the information in the operating instructions Please read these operating instructions carefully before commissioning, even if you already have experience with SAUTER measuring instruments. Staff training The device may only be operated and maintained by trained staff. Note: To view the CE declaration, please click on the following link: https://www.kern-sohn.com/shop/de/DOWNLOADS/...