Standard 12-Lead - Schiller CARDIOVIT CS-104 User Manual

Table of Contents

Advertisement

4
Recording an ECG
4.1
Placing the Electrodes
C1 Red
R Red
N Black
IEC Label
C1 white / red
C2 white / yellow
C3 white / green
C4 white / brown
C5 white / black
C6 white / violet
L yellow
R red
F green
N black
Page 48
4.1.3

Standard 12-lead

AHA Label
V1 brown / red
V2 brown / yellow
V3 brown / green
V4 brown / blue
V5 brown / orange
V6 brown / violet
LA black
RA white
LL red
RL green
• Auto Interpretation is only generated when standard 12-lead electrode lead config-
uration is set.
• The lead configuration is set in the lead configuration screen selected at the start
of the recording.
• The electrode resistance is constantly monitored in the recording screen and a
lead-off indication displayed if the resistance is too high.
• When making an ECG with a child it is sometimes physically difficult to position all
electrodes. When this is the case electrode C4 can be placed on the right side of
the chest.
Electrode Placement
 Fourth intercostal space at the right sternal border
 Fourth intercostal space at the left sternal border
 Midway between C2 and C4
 Left mid-clavicular line in the fifth intercostal space
 Left anterior axillary line on the same horizontal level as C4
 Left mid-axillary line on the same horizontal level as C4
 Left arm
 Right arm
 Left foot
 Right foot
CARDIOVIT CS-104