4

Recording an ECG

4.1

Placing the Electrodes

Page 46
4 Recording an ECG
 Ensure that neither the patient nor the leading parts of the patient connection nor
the electrodes (including the neutral electrodes) come in contact with other
persons or conductive objects, even when these are earthed.
4.1
Placing the Electrodes
• Electrode placement assistance and electrode check is also given in the ECG ac-
quisition screen
page 60).
• The IEC or AHA lead cable is set in ECG settings
132).
4.1.1

Electrode Identification and Colour Code

The electrode placements shown in this section are labelled with the colours
according to Code 1 (IEC) requirements. The equivalent Code 2 (AHA) colours are
given below.
Limb
Chest
according
to Wilson
Neutral
(see para. 4.3, Hookup Screen, Electrode Placement and Check,
Code 1 (IEC)
IEC Label
Colour
R
Red
L
Yellow
F
Green
C1
White / Red
C2
White / Yellow
C3
White / Green
C4
White / Brown
C5
White / Black
C6
White / Violet
N
Black
CARDIOVIT CS-104
(see para. 11.11, ECG, page
Code 2 (AHA)
AHA Label
Colour
RA
White
LA
Black
LL
Red
V1
Brown / Red
V2
Brown / Yellow
V3
Brown / Green
V4
Brown / Blue
V5
Brown / Orange
V6
Brown / Violet
RL
Green