Automatic Resting Ecg Recording - Schiller CARDIOVIT AT-1 G2 User Manual

Table of Contents

Advertisement

CARDIOVIT AT-1 G2
Auto
Review
The recording can be reviewed.
Use the keys  to toggle
between the following 6 pages:
• Rhythms I, II, III
• Rhythms aVL, aVL, -aVR
• Rhythms V1, V2, V3
• Rhythms V4, V5, V6
• Interpretation
• Measurements
Colour code for classification:
Blue = undefined
Red = abnormal ECG
Orange = otherwise normal ECG (or)
possibly abnormal ECG
User guide
6.2

Automatic resting ECG recording

To take an automatic ECG recording, press the Auto key. After approx. 10 seconds,
the recording is analysed and the result displayed. The recording can be checked and
printed.
12345
60
/min
I
II
III
Normal ECG
HR
60 /min
Sinus rhythm
Normal electrical axis
Non specific ST abnormality (elevation)
next/prev. page
 Use the keys
 Print the recording
(Depending on the settings, not all pages are printed as shown on the display.)
 Press
deleted.
• When the ECG has been printed, the patient data is deleted; however, the patient
data is again activated if the key
Automatic resting ECG recording
25 mm/s 10 mm/mV
RR
1000 ms
QRS
P
116 ms
QT
PQ
172 ms
QT
c
25 mm/s 10 mm/mV
accept & print
to review the recording, pages 1 - 6

and check/complete the patient data on the printout.
to exit the review screen without printing. The patient data is not
is pressed twice.
Resting ECG
28.08.2016 17.58
LP 150Hz AC 50 Hz
14.03.2017 17:58
14.03.2017
94 ms
P
46°
416 ms
QRS 47°
416 ms
T
36°
LP 150Hz AC 50 Hz
discard
6
6.2
17:45
Page 31

Advertisement

Table of Contents
loading