Meter Error Codes - Veris Industries FSR Series Installation Manual

Ultrasonic flow meter
Table of Contents

Advertisement

TM
Error Codes
(revised 5-25-2009)
Code Number
0001
Serial number not present
0010
Signal Strength is below Signal Strength Cutoff entry
Measured Speed of Sound in the liquid is greater than ±10% different
0011
than the value entered during meter setup
1001
System tables have changed
1002
System configuration has changed
3001
Invalid hardware configuration
3002
Invalid system configuration
3003
Invalid strategy file
3004
Invalid calibration data
3005
Invalid speed of sound calibration data
3006
Bad system tables
4001
Flash memory full
Table A-5.1 - Error Codes
Z205739-0D
PAGE 46
Alta Labs, Enercept, Enspector, Hawkeye, Trustat, Aerospond, Veris, and the Veris 'V' logo are trademarks or registered trademarks of Veris Industries, L.L.C. in the USA and/or other countries.
FSRxxxx SERIES
Description
Class C Errors
Class B Errors
Class A Errors
©2013 Veris Industries
Warnings
Hardware serial number has become inoperative – system performance will not be
influenced.
Low signal strength is typically caused by one of the following:
» Empty pipe
» Improper programming/incorrect values
» Improper transducer spacing
» Non-homogeneous pipe wall
Verify that the correct liquid was selected in the BASIC menu.
Verify that pipe size parameters are correct
Initiate a meter RESET by cycling power or by selecting SYSTEM RESET in the SEC
MENU.
Initiate a meter RESET by cycling power or by selecting SYSTEM RESET in the SEC
MENU.
Upload corrected file
Upload corrected file
Upload corrected file
Re-calibrate the system
Upload new data
Upload new table data
Return unit to factory for evaluation
INSTALLATION GUIDE
Correction