Sen Menu -- Sensor Menu; Sec Menu -- Security Menu - Veris Industries FSR Series Installation Manual

Ultrasonic flow meter
Table of Contents

Advertisement

TM
3. Outputs
Two independent open collector transistor outputs are available. Each output can be
configured independently for one of the following.
CONTROL/HZ -- Control Options (Choice): Select either Control 1 or Control 2 to
program
(A) TOTALIZE -- Totalizer Output Options
• TOT MULT --Totalizer Multiplier (Value) - Sets the multiplier value applied
to the totalizing pulse output
(B) FLOW -- Flow Alarm Output Options: FLOW -- Flow Alarm Values
• ON (Value): Sets value at which the alarm output turns on
• OFF (Value): Sets value at which the alarm output turns off
(C) SIG STR -- Signal Strength Alarm Options: SIG STR -- Signal Strength
Alarm Values:
• ON (Value): Sets value at which the alarm output turns on
• OFF (Value): Sets value at which the alarm output turns off
(D) ERRORS: Alarm outputs on any error condition. See Error Table in the
Appendix of this manual.
(E) NONE: Alarm outputs disabled
Note: The setup options for both CONTROL 1 and CONTROL 2 follow the same menu path. For a
complete view of the menu options, see the Menu Map in the Appendix of this manual.
Z205739-0D
PAGE 24
Alta Labs, Enercept, Enspector, Hawkeye, Trustat, Aerospond, Veris, and the Veris 'V' logo are trademarks or registered trademarks of Veris Industries, L.L.C. in the USA and/or other countries.
FSRxxxx SERIES
©2013 Veris Industries
INSTALLATION GUIDE

SEN Menu -- Sensor Menu

The SEN MENU allows access to the various types of transducers the meter can work
with. Selecting the proper transducers in conjunction with the transducer mount
(XDCR MNT) and transducer frequency (XDCR HZ) is critical to accurate operation of
the meter.
SEN MENU -- Transducer Selection Menu (Choice)
FST4, FST5: Used on pipes 2 inches (51 mm) and larger. (250°F/121°C
maximum)
COPPER PIPE: Used with FST1, FST2, and FST3 small pipe transducers.
185°F/85°C maximum
ANSI PIPE: Used with FST1, FST2, and FST3 small pipe transducers. 185°F/85°C
maximum
TUBING: Used with FST1, FST2, and FST3 small pipe transducers. 185°F/85°C
maximum

SEC Menu -- Security Menu

The SEC MENU menu allows access to meter functions that may need to be protected
from changes.
SEC MENU -- Security Function Selection Menu
TOT RES -- Totalizer Reset (Choice)
• YES
• NO
Resets the totalizing displayed on the LCD to zero
SYS RES -- System Reset (Choice)
• YES
• NO
Restarts the flow meter's microprocessor. This is similar to power cycling
the flow meter.
CH PSWD? -- Change Password (Value): 0 - 9999
The password comes from the factory set to 0000. When set to 0000 the password
function is disabled. Changing the password to any other value from 0001 to 9999
enables this function, and configuration parameters are inaccessible unitl the
password is entered. If the password is lost or forgotten, contact the manufacturer
for a universal password to unlock the meter.