6. Close the battery cover of the data logger. Push the latch down until you hear a click. If
the cover does not close easily, push the probe (or extension cable) in and try again.
7. Look at the display and wait for the data logger to start up. Then verify the status:
• Data logger shows the word NEW on each affected channel when it detects that a
probe has been changed
display shows the measurement reading(s).
Figure 1 Detection of a new RH + T probe
• If dashes "- - -" are shown instead of measurement readings, check that the probe is
properly connected. It is possible to disconnect the probe or cable by pulling it with
the battery cover open.
• Battery indicator should show full batteries
• There should be no error codes shown. If there are, see
1)
This feature is included in RFL100 firmware version 1.2.0 and newer.
6
1)
. The text is shown for a few seconds, after which the
.
Error codes (page
19).
M211822EN-F