Download Print this page

DentalEZ StarDental Titan 3 Five Star Instruction Manual page 2

Lubricated prophy angle non-sterile

Advertisement

Cleaning
7. When prophy paste is consistently found in the cartridge, install a new
bronze bushing, or a new cartridge, StarDental Part #258948, which
Manual
includes the bronze bushing. Replacing the bronze bushing renews the
Continued:
seal and protects the gear and bearing from abrasive paste.
8. Wipe the inside of the housing with a lint-free
cloth or cotton swab.
Note: Do not immerse head in solvent.
9. After cleaning, apply one drop of DentaLube
II, StarDental Part #262539, to the inside
diameter of the O-ring and to the gear as
shown. (See Illustration to the right)
10. To reassemble, insert gear assembly into bronze bushing and guide shaft
through the hole in O-ring. Install cartridge into prophy head angle.
11. Turn end cap counterclockwise and hand tighten to secure. Use wrench
to fully seat – 30° rotation maximum.
NOTICE
Do not overtighten as poor performance and damage to angle may result
Lubricate - See 'Lubrication' instruction section.
Lubrication:
Disinfection:
Use steam sterilization exclusively for complete sterility as outlined below.
Angle should be routinely maintained, checked for corrosion, worn parts,
Maintenance:
etc. and should be replaced on an as needed basis.
Inspection &
See Installation, Operating Recommendations, and/or Troubleshooting sections of
function testing:
this manual.
Single: A standard sterilization packaging material must be used. Ensure that the
Packaging:
pack is large enough to contain the item without stressing the seals.
Gravity Type Autoclave:
Wrapped
Instrument
• Minimum of 15 minutes at 132° C
Sterilization:
• Minimum of 30 minutes at 121° C
Do not exceed 135° C
Drying:
Dry the prophy angle for 30 minutes minimum using the sterilization dry cycle and
allow the prophy angle to cool to room temperature in bag prior to use. If moisture is
present, increase your sterilizer dry cycle appropriately.
Storage:
Store in bag after sterilization until ready to use.
Do not exceed the sterilizer manufacturer's guidelines for load capacity. Use only
Additional
paper or paper/plastic bags.
Information:
See first or last panels of this manual.
Manufacturer
Contact:
Troubleshooting
Problem: LOW POWER
Cause: Air leak at motor
Correction: Replace gasket;
refer to Titan® Motor Instruction Manual.
Cause: Debris build-up
Correction: Refer to Cleaning and Lubricating section.
Cause: Faulty angle
Correction: Replace cartridge.
Note:
If a problem persists, contact dealer for further assistance;
or, contact StarDental® technical service at: 1-866-DTE-INFO (1-866-383-4636)
Cartridge Replacement
1. Remove the end cap using the plastic wrench to turn it clockwise. (See illustration
under Cleaning and Lubricating.)
Note: Do not use pliers on the bronze bushing.
2. Pull bronze bushing to remove cartridge from prophy head as shown below. If
cartridge is not removable by hand, thread a prophy cup or a screw-to-button
adapter in the cartridge and tug on it.
3. Separate gear assembly from the bronze bushing by pulling apart as shown above.
4. If gears are worn, replace cartridge, StarDental Part #258948. If not, simply wipe
clean with a lint free cloth.
5. Wipe the inside of the housing with a lint free cloth or cotton swab, barely dampened
with solvent.
Pre-Vacuum Autoclave:
Note:
Do not immerse head in solvent.
• Minimum of 4 minutes at 132° C
6. Guide the shaft of the gear assembly through the center of the bronze bushing.
Install the assembled cartridge into the angle housing.
7. Position end cap and turn counterclockwise. Hand tighten to secure. Use wrench
to fully seat – 30° maximum.
Do not overtighten as poor performance and damage to angle may result.
NOTICE
Replacement Part Numbers
Always insist on StarDental Parts for best performance.
Description
Titan
®
3 Five Star Prophy Angle ......................................................................................... 265803
Titan 3 Five Star Prophy Angle Cartridge Lubricated ................................................ 258948
(includes cartridge and bronze bushing with O-ring)
Transmission Gear Assembly Lubricated ......................................................................... 259008
Prophy Angle End Caps (Package of 2) Lubricated ...................................................... 265805
Screw-to-Button Adapter ........................................................................................................ 258950
End Cap Wrench (Plastic) ....................................................................................................... 258953
Angle Disconnect Wrench ...................................................................................................... 203048
O-Rings (Package of 3) ..............................................................................................................258951
DentaLube II .................................................................................................................................262539
Titan 3 Motor-to-Angle Adapter...........................................................................................264073
Titan T Motor-to-Angle Adapter ..........................................................................................263970
Cartridge
Bronze
Bushing
Gear
Assembly
Item Number
Important Notice About Returning Product To The DentalEZ Inc.
Anyone wishing to return any DentalEZ Inc. product to a DentalEZ facility MUST obtain the proper
RETURN AUTHORIZATION NUMBER from the Customer Service Department. The RA Number must
appear clearly marked on both the outside and inside of the returned product carton. When calling
for a return authorization number you will be asked to provide the following information:
•Product Name, Model Number or Part Number
To Obtain A
•Serial Number of Product
Return Authorization
•Dealer Return Purchase Order Number
Number Call
•Reason For Return
1-866-DTE-INFO
•Copy of Invoice Showing Purchase
Items will be evaluated. Return Authorization Number is not a guarantee for credit. Your cooperation
with this procedure will help us expedite processing of any returns.
For new products, manuals and technical information call
1-866-DTE-INFO or visit our web site @ www.dentalez.com
Limited Warranty
O-Rings and Seals are specifically excluded from this warranty.
DentalEZ ® warrants the Titan ® Lubricated Five Star Prophy Angle to be free of
defects in material and workmanship, under normal usage, under the following terms:
StarDental® Products:
Lubricated Five Star Prophy Angle
Please note the following additional terms of our warranty and return policy:
• Warranties cover manufacturing defects only and do not cover defects resulting
from abuse, improper handling, cleaning, care or maintenance, normal wear and
tear or non-observance of operating, maintenance or installation instructions.
Failure to use authorized parts or an authorized repair facility voids this warranty.
• Liability is limited to repair or replacement of the defective product at our sole
discretion. All other liabilities, in particular liability for damages, including,
without limitation, consequential or incidental damages are excluded.
• THIS WARRANTY IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESSED OR
IMPLIED, INCLUDING ANY IMPLIED WARRANTIES OF MERCHANTABILITY OR
FITNESS FOR A PARTICULAR PURPOSE. NO EMPLOYEE, REPRESENTATIVE
OR DEALER IS AUTHORIZED TO CHANGE THIS WARRANTY IN ANY WAY OR
TO GRANT ANY OTHER WARRANTY.
WARRANTY REPAIRS: Parts repaired or replaced on a product that is in warranty
will be warranted for the duration of that product's original warranty.
NON-WARRANTY REPAIRS: The warranty on parts either repaired or replaced on
an out-of-warranty product will cover the repaired part only and will be for the
time frame of a new parts warranty period.
PRODUCT RETURN: Opened products or product returns more than a year old
cannot be returned for credit. There will be a 15% ($25.00 minimum) restocking
charge on all items authorized for return.
* Provided conditions in warranty are met.
WARRANTY REGISTRATION
PLEASE VISIT THE WEBSITE BELOW TO REGISTER yOUR STARDENTAL PRODUCT. THIS WILL ENSURE
PROPER REGISTRATION AND PROMPT AND ACCURATE SERVICE SHOULD ANy OF yOUR PURCHASES
NEED EVALUATION OR REPAIR DURING THE WARRANTy PERIOD.
StarDental® Products:
www.dentalez.com/warranty
Warranty Period*:
3 Months from date of purchase

Advertisement

loading

This manual is also suitable for:

265803