DentalEZ everLight EVLCE Installation, Operation & Care Manual

DentalEZ everLight EVLCE Installation, Operation & Care Manual

Track, ceiling, wall & cabinet mount

Advertisement

Quick Links

®
Track, Ceiling, Wall & Cabinet Mount
DentalEZ MAKES YOUR PRACTICE PERFECT
®
®

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the everLight EVLCE and is the answer not in the manual?

Questions and answers

Summary of Contents for DentalEZ everLight EVLCE

  • Page 1 ® Track, Ceiling, Wall & Cabinet Mount DentalEZ MAKES YOUR PRACTICE PERFECT ® ®...
  • Page 2: Table Of Contents

    Table of Contents Section I Introduction Specifications ......................... 2 Classifications ........................2 Dimensions ........................3-4 Light Head Range of Motion ....................5 Packaging ..........................5 Section II Installation Track Mount Assembly ....................6-10 Ceiling Mount Assembly ..................... 11-14 Wall Mount Assembly ....................15-16 Cabinet Mount Assembly ...................
  • Page 3: Section I Introduction

    Section I Introduction EverLight ® This manual contains installation, operation and care –– Type of protection against electric shock: Class 1 instructions and user service information for the Equipment EverLight ® LED light. –– Degree of protection against the ingress of water: Ordinary The EverLight is designed to provide trouble-free service when installed, operated and maintained...
  • Page 4: Dimensions

    Section I Introduction Dimensions 24" 1.61" 1.25" 3.5" 9" 14.375" 15" 17" Track Mount 1.375" 24" 7.25" 31.875" 9" 14.375" 15" 17" Ceiling Mount EverLight ®...
  • Page 5 Section I Introduction EverLight ® 24" 1.375" 4.5" 9" 56" 14.375" 15" 17" Wall/Cabinet Mount Installation, Operation and Care Manual...
  • Page 6: Light Head Range Of Motion

    All parts supplied are vital for proper installation. Do not discard any hardware or components — NOTICE — until installation is checked and complete. For any questions about an order, please contact a DentalEZ Equipment customer service representative at 1-866-DTE-INFO (1-866-383-4636). EverLight ®...
  • Page 7: Section Ii Installation

    Section II Installation EverLight ® 1. After determining the location of the track Track Mount Assembly light (see illustration below), arrange for the hole for the power cord to be above the head Tools Required of the chair. •• Phillips-head Screwdriver ••...
  • Page 8 Section II Installation NOTE: Refer to the exploded illustration below Side during track light installation. Cover 5. Loosen, but do not remove, two Phillips- head screws from one end cap of the track mount assembly. End Cap 6. After loosening the end cap (do not remove end cap), take off the side covers and set Phillips-head them aside.
  • Page 9 Section II Installation EverLight ® 7. Remove the four screws from the power Power Supply Wiring supply housing cover, then take off the cover 1. Make sure the appropriate electrical line and set the screws and cover aside. (100-240 VAC, 50/60 Hz) is properly Power Supply supplied to the power supply housing.
  • Page 10 Section II Installation NOTE: This hole is for the 5/16" bolt that secures 4. Tuck the wires in neatly and replace the power the drop post to the support hub. supply housing cover using the retaining screws previously removed. 4. Route the pigtail from the trolley through the drop post 5.
  • Page 11 Section II Installation EverLight ® Flex Arm Height Adjustment 1. An adjustment can be made to the flex arm travel in the event the flex arm contacts the track or the ceiling. Drop Post 2. To access the adjustable stop, remove the four screws that hold the arm cover.
  • Page 12: Ceiling Mount Assembly

    — NOTICE — individual preferences. But, be aware of the following For any questions about an order, please contact framing considerations. a DentalEZ Equipment customer service repre- •• Check ceiling structure and assure it will sentative at 1-866-DTE-INFO. support a minimum of 200 pounds.
  • Page 13 Section II Installation EverLight ® Power Supply Wiring •• Ensure no lateral movement of the light column is possible. 1. Make sure the appropriate electrical line •• Check for an electrical circuit near the (100-240 VAC, 50/60 Hz) is properly support area of the light.
  • Page 14 Section II Installation Drop Post Installation 5. Insert the cut end of the drop post into the support hub and tighten the 5/16" bolt NOTE: The ceiling light is designed to accommo- through the drop post and let the post rest on date up to an 10' ceiling.
  • Page 15 Section II Installation EverLight ® Flex Arm Height Adjustment 8. Slide the cover and cover support up the drop post and tighten the set screw in the cover An adjustment can be made to the flex arm to limit support to prevent the cover from falling. the vertical travel of the arm.
  • Page 16: Wall Mount Assembly

    Section II Installation a. First identify the center point of the long Wall Mount Assembly arm rotation. This point should align with the chair's center line and off-set 25-1/4" Tools Required from the front edge of the chair's base plate in the direction of the headrest. ••...
  • Page 17 Section II Installation EverLight ® Black — NOTICE — White Wire Harness Wire All wiring must be in accordance with local From Light electrical codes. Green/Yellow Mounting Top Hole for Wire Bracket Wall Mount Light Arm Power Cover Supply Screws Harness 10.
  • Page 18: Cabinet Mount Assembly

    Magnetic Level heads, smoke alarms, etc.! •• Lubricant: Silicon/Grease Product The illustration below gives the suggested mounting location for the DentalEZ cabinet mount assembly Power Supply Wiring and the HDX Intraoral X-ray. Make sure the required power supply wiring is...
  • Page 19 EverLight ® NOTE: 1. Secure the vertical mounting post(s) to the The metal post support brackets are metal frame inside the cabinet assembly using designed to facilitate per pendicular and the metal post support brackets supplied. parallel adjustments. (Refer to the illustration below.) 3.
  • Page 20: Section Iii Operation

    Section III Operation Procedural Settings The EverLight dental light is designed to provide illumination of the oral cavity for use by dental •• A - Sun - Color Matching - 5300° Kelvin professionals in the dental operatory. •• B - Sun/Bulb - Operative Setting - 4300° Kelvin Light Switch ••...
  • Page 21: Positioning Light

    Section III Operation EverLight ® Positioning Light Using the handles on either side of the LED head, position the light to illuminate the oral cavity. (See Light Head Range Of Motion, Page 5.) Handles — HELPFUL NOTE — To take full advantage of the EverLight's intensity and color correction, position the light between 20"...
  • Page 22: Eliminating Adjustable Arm Drift (Wall Mount)

    Section III Operation 3. Insert a 9/64 Allen wrench into the two Eliminating Adjustable Arm Drift small holes in the side of the adjustable arm and completely loosen the friction block. The wall mount's adjustable arm is factory-adjusted to counter-balance the weight of the light head. If it becomes necessary to adjust the mechanism because 9/64 Allen of added weight, such as a mirror, perform the...
  • Page 23: Section Iv Care

    Section IV Care EverLight ® Cleaning & Disinfecting — CAUTION — Avoid spraying anything directly on the light. Spray cloth first, then wipe. 1. To clean all surfaces, use a typical ALCOHOL-FREE, commercial cleaner or a solution of mild liquid detergent and water. 2.
  • Page 24: Section V User Service Information

    Troubleshooting If the area of concern is not addressed in this The EverLight is designed to provide trouble-free manual, contact your local DentalEZ full-service service when installed, operated and maintained dealership. (See Product Support Services, Page 25.) according to the procedures set forth in this manual.
  • Page 25: Section Vi Parts List/Diagrams

    Section VI Parts List/Diagrams EverLight ® (Wall light arm assembly shown) Light Arm Light Head c t i & c t i y t i , t i Drop Post (Drop Post for Ceiling and Track mount only) & l i e Installation, Operation and Care Manual...
  • Page 26: Product Support Services

    EverLight ® DentalEZ Group and its employees are proud of the products we provide to the dental community. We stand behind these products with a warranty against defects in material and workmanship as provided below. In the event that you experience difficulty with the application or operation of any of our products, please contact our Technical Service Department at our expense at 1-866-DTE-INFO (1-866-383-4636).
  • Page 27: Wiring Schematic

    Wiring Schematic EverLight ® OUTPUT INPUT 12 VDC, 40 W 100-240 VAC, 50/60 HZ Power Supply Black (Line) White (Neutral) Green (Ground) Installation, Operation and Care Manual...
  • Page 28: Emc Information

    EMC Information — WARNING — Use only replacement cables listed in Parts Section, Page 24. Other ● cables and accessories may negatively affect EMC performance. When the Everlight is used adjacent to other equipment, observe the ● light to verify normal operation. Guidance and manufacturer’s declaration - electromagnetic emissions The Everlight Model EVLCE is intended for use in the electromagnetic environment specified below.
  • Page 29 EMC Information EverLight ® Guidance and manufacturer’s declaration – electromagnetic immunity The Everlight Model EVLCE is intended for use in the electromagnet environment specified below. The customer or the end user of the Model EVLCE should assure that it is used in such an environment.
  • Page 30 Ceiling Mount Light Full-Size Template PN: 2677-069A...
  • Page 31 3/4" 16" 8" Ø1-1/2" Hole for Wiring Through Wall 12" 10" 6-3/4" Drill Hole Size Depends on Size of Wall Fastener Chosen (4 Places) 8-3/4" 1" 17-1/2" Wall Mount Light Full-Size Template PN: 2677-070A...
  • Page 32 2500 Highway 31 South Bay Minette, Alabama 36507 1-866-DTE-INFO Fax: (251) 937-0461 www.dentalez.com © DentalEZ Alabama, Inc. Printed in USA PN: 2717-241B March, 2011...

Table of Contents