3.1.2
Electrical Requirements
Verify the system is to be operated in a room that meets the electrical requirements listed below.
Table 3-3: Electrical Specifications
Electrical Rating
Measurement (System Only)
Current Draw (System Only)
Power Draw (System Only @ peak)
Leakage Current (System Only)
Note: Refer to OEM peripheral manuals for peripheral-specific electrical ratings.
3.1.2.1 Instrument Input Power Rating
An instrument input power rating has been established to clarify the maximum power
requirement of the system with all accessories. Depending upon the accessories and options
installed, the system may not draw the full amount of power listed below:
Table 3-4: Instrument Power Input Rating
System
Voltage, VAC
115V
220V
For optimal system performance, use a dedicated, interference-free, isolated, grounded wall
outlet. To ensure grounding reliability, use a hospital-grade power cord and connect it only to an
equivalent hospital-grade socket. The specifications of the hospital-grade power cord as follows:
Table 3-5: Hospital-Grade Power Cord Specifications
Input Voltage
100-120V ~, 47-63Hz
200-240V~, 47-63Hz
Chapter 3: System Installation
Value
115 Vac @ 47-63Hz (Canada and USA)
(International)
4.00A (Canada and USA)
(International)
450 VA
< 100 uA
Nominal System Power Usage,
including isolated accessory power
7A Continuous
4A Continuous
Hospital Grade Power Cord Specifications
125Vac, 15A, 3 wire, 18 AWG, grounding type, 5-15P Hospital Grade plug
cap, less than 6 m long, CSA & UL approved
250Vac, 15A, 3 wire, 18 AWG, grounding type, 6-15P Hospital Grade plug
cap, less than 6 m long, CSA & UL approved
10
SONIX System
Isolated Accessory
Power Usage
Power Available
5.5A Continuous
1.5A Continuous
3.25A Continuous
0.75A Continuous
00.053.009, Revision A SONIX MDP Service Manual